文献速递临床生化质量管理与危急值

Topper系统自采毛细血管血用于15种常规临床化学分析物的验证

Validation of (self-collected) capillary blood using a topper collection system as alternative for venous sampling for 15 common clinical chemistry analytes.
PubMed 文献 · Clin Chem Lab Med · 2025-06-02

研究纳入120名参与者(21名健康志愿者和99名患者),比较Topper系统自采毛细血管血与静脉血在15种常规临床化学分析物上的一致性。采用Spearman相关、Passing-Bablok回归和Bland-Altman差异分析,以基于生物学变异(EFLM数据库)的允许偏差标准判定可比性。结果显示113/120(94%)自采样本可分析至少一种分析物,多数分析物偏差在可接受范围内。

要点

  • 94%自采样本可成功分析至少一种分析物
  • 葡萄糖和未校正AST未达到最小偏差标准
  • 肌酐和白蛋白在最小偏差标准内,尿素和钙在理想偏差标准内

核心结论

  • 自采毛细血管血可作为大多数(若非全部)所研究分析物静脉采血的可替代方案。

研究设计与样本

  • 研究纳入120名参与者(21名健康志愿者和99名患者),先采集静脉样本,再由参与者使用Topper系统自采毛细血管血,血清在Cobas Pro(Roche)系统上分析15种常规临床化学分析物及血清指数,采用Spearman相关、Passing-Bablok回归和Bland-Altman差异分析,以基于生物学变异(EFLM数据库)的允许偏差标准判定可比性。

主要发现

  • 113/120(94%)自采样本可分析至少一种分析物。
  • Bland-Altman差异分析显示葡萄糖和未校正AST未达到最小偏差标准。
  • 肌酐和白蛋白在最小偏差标准内,尿素和钙在理想偏差标准内。
  • ALP、校正AST、ALT、总胆固醇、HDL胆固醇、LDL胆固醇、CRP、GGT、总蛋白和甘油三酯均在最佳偏差标准内。

关键数值与推荐

  • 可比性判定采用基于生物学变异(EFLM数据库)的允许偏差标准。

临床怎么用

  • 基于分析物在样本处理前的稳定性,家庭采样对于部分分析物似乎是可靠的采样选择。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Home blood sampling for clinical purposes has gained much interest. This study validates the recently developed Topper-based capillary blood collection procedure for 15 commonly used chemistry analytes. METHODS: A total of 120 study participants (21 healthy volunteers and 99 patients) were enrolled. A venous sample was obtained and then participants were asked to self-collect blood by the Topper system (SelfSafeSure Blood Collection Devices). Collected sera were analyzed for 15 common clinical chemistry analytes and the serum indices on a Cobas Pro (Roche) system. Spearman correlations, Passing-Bablok regression analysis, and Bland-Altman difference analysis were performed. Comparability was determined using allowable bias criteria based on biological variation (EFLM database). RESULTS: In 113 out of 120 (94 %) self-collections a sample was obtained that allowed for the analysis of at least one analyte. Bland-Altman difference analysis showed that glucose and uncorrected AST did not meet the minimum bias criterion, creatinine and albumin were within the minimum bias criterion and urea and calcium were within the desirable bias criterion. ALP, corrected AST, ALT, total- HDL- and LDL-cholesterol, CRP, GGT, total protein and triglycerides were all within the optimal bias criterion. CONCLUSIONS: Our study demonstrates that self-collected capillary blood can be used as a reliable alternative to venous sampling for most, if not all, analytes studied. Based on the analyte stability prior to sample processing, home sampling appears to be a reliable sampling option for a selection of these analytes.

来源:Clin Chem Lab Med,查看原文。版权归原作者所有。

引用

  1. Geujar R, Treffers D, Roelofs M, et al. Validation of (self-collected) capillary blood using a topper collection system as alternative for venous sampling for 15 common clinical chemistry analytes. Clin Chem Lab Med. 2025. PMID: 40459334. doi:10.1515/cclm-2025-0148

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。