文献速递临床生化质量管理与危急值

单克隆蛋白导致溶血指数假性升高的钾检测新陷阱

A new pitfall in potassium testing: A case series of false elevation of hemolysis index in the presence of monoclonal protein.
PubMed 文献 · Clin Chim Acta · 2025-03-05

本病例系列报告了单克隆蛋白存在时溶血指数假性升高,导致实验室信息系统算法不适当地抑制钾结果。样本在Atellica CH930自动生化分析仪上被溶血指数判定为强溶血,但视觉检查显示血浆清晰,提示溶血指数升高并非由溶血引起。报告强调需警惕单克隆蛋白对溶血指数的潜在干扰,并提出三个警示信号以识别此类异常。

要点

  • 单克隆蛋白可导致溶血指数假性升高,进而使LIS算法错误抑制钾结果。
  • 溶血指数升高但离心后视觉检查血浆清晰时,应怀疑单克隆蛋白干扰。
  • 报告提出三个警示信号帮助识别溶血指数假性升高。

核心结论

  • 单克隆蛋白可干扰溶血指数,导致钾结果被实验室信息系统算法不适当地抑制。

研究设计与样本

  • 病例系列报告,样本在Atellica CH930自动生化分析仪上检测溶血指数,并进行离心后视觉检查。

主要发现

  • 样本被溶血指数判定为强溶血,但视觉检查显示血浆清晰,提示溶血指数升高并非由溶血引起。
  • 溶血指数假性升高导致实验室信息系统算法不适当地抑制钾结果。

临床怎么用

  • 当溶血指数升高但离心后视觉检查血浆清晰时,应警惕单克隆蛋白干扰,避免误判为溶血。
  • 报告提出三个警示信号用于检测此类异常,以更好地解释结果。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Hemolysis is the most common preanalytical interfering factor and a major source of error in clinical biochemistry analysis. Modern laboratories routinely assess hemolysis in conjunction with a laboratory information system algorithm that automatically adds a comment or rejects test values when appropriate. The hemolysis index has largely replaced visual inspection for assessing specimen quality prior to clinical chemistry testing. However, visual inspection of samples after centrifugation can provide valuable information about preanalytical interferences. In this report, we present a case series of inaccurate assessment of hemolysis index associated with the presence of monoclonal protein leading to inappropriate suppression of potassium results by the laboratory information system algorithm. These samples were originally identified as strongly hemolysed by the hemolysis index on the Atellica CH930 automated chemistry analyser. However, visual inspection revealed clear plasma, suggesting that the elevated hemolysis index was not due to hemolysis. This report highlights the need to be aware of potential monoclonal protein interference with the hemolysis index for better interpretation of results, and proposes three warning signs to detect this type of anomaly.

来源:Clin Chim Acta,查看原文。版权归原作者所有。