双模块比值作为基于患者数据的实时质控指标用于小偏移检测与比对

Double-module ratio as a patient-based real-time quality control index for small shift detection and comparative assay.
PubMed 文献 · Clin Chim Acta · 2025-02-24

研究纳入同部门集中检测的常规实验室数据,将两个模块的PBRTQC数据比值定义为双模块比值(RAlb、RCa、RChE、RTP),并与常规PBRTQC算法比较。四种比值识别系统误差的能力均优于对应单模块指标,其中RAlb、RCa、RTP可显著辅助检测约小于允许总误差1.5倍的小系统误差,RChE对负向系统误差的监测能力也较ChE明显改善。

要点

  • 双模块比值识别系统误差的能力优于对应单模块PBRTQC指标。
  • RAlb、RCa、RTP可显著辅助检测约小于允许总误差1.5倍的小系统误差。
  • 比值对随机误差检测的影响取决于指标和PBRTQC参数。

核心结论

  • 双模块比值作为PBRTQC指标可促进小偏移检测,并实时确保多模块间结果一致性。

研究设计与样本

  • 纳入同部门集中检测、原始患者数据在日内不规则波动的实验室结果。
  • 将两个模块PBRTQC数据的比值定义为双模块比值(RAlb、RCa、RChE、RTP)。
  • 采用MA、移动中位数、指数加权MA和移动标准差等PBRTQC程序比较四对指标。

主要发现

  • 四种比值(RAlb、RCa、RChE、RTP)识别系统误差的能力均强于对应的Alb、Ca、ChE、TP。
  • 改善程度与取比值后QC数据变异的降低程度成比例。
  • RAlb、RCa、RTP可显著辅助检测约小于允许总误差1.5倍的小系统误差。
  • ChE对负向系统误差几乎无监测能力,而RChE有显著改善。
  • 比值对随机误差检测的影响取决于指标和PBRTQC参数。

临床怎么用

  • 可将双模块比值作为PBRTQC指标,用于小系统误差检测和多模块结果一致性实时监控。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Irregular fluctuations in patient data affect patient-based real-time quality control (PBRTQC) performance. Moreover, results among multiple instruments can be inconsistent. Here, we explored a simple index, double-module ratio, to improve PBRTQC performance through overcoming irregular fluctuations, and ensure consistency across multiple instruments in real time. METHODS: Laboratory results from the same departments tested centrally such that the original patient data fluctuated irregularly during the day were included. Albumin (Alb), calcium (Ca), cholinesterase (ChE), and total protein (TP) indexes derived from a single module were processed through conventional PBRTQC. The ratios calculated from PBRTQC data of two modules were defined as double-module ratios (RAlb, RCa, RChE, and RTP). The four index pairs were compared using various PBRTQC procedures, including conventional algorithms: moving average (MA), moving median, exponentially weighted MA, and moving standard deviation. RESULTS: All four ratios (RAlb, RCa, RChE, and RTP) were more powerful in identifying system error (SE) than their counterparts (Alb, Ca, ChE, and TP). These improvements were proportional to the degree of reduction in QC data variation after ratios were taken. In particular, RAlb, RCa, and RTP could aid in detecting small SEs (about <1.5 folds of the allowable total errors) significantly. Although ChE demonstrated almost no surveillance capacity for negative SEs, RChE demonstrated a significant improvement. The impact of the ratios on random error detection was dependent on the indexes and PBRTQC parameters. CONCLUSIONS: As a PBRTQC index, double-module ratios can facilitate small shift detection and ensure result consistency among various modules in real time.