血清25(OH)D LC-MS/MS检测的分析性能评价与优化
本研究评估中国12家临床实验室现行LC-MS/MS法检测血清25(OH)D的性能,并与优化方法比较。优化法使更多实验室达到精密度和正确度标准,测量不确定度降低,但低浓度样本仍难达标。
要点
- 现行方法仅43%和57%实验室达到CV<12.5%和平均偏倚<8.3%标准。
- 优化法显著降低测量不确定度,但低浓度样本仍难达MU<13.6%。
核心结论
- 应用优化LC-MS/MS方法后,血清25(OH)D检测的精密度、正确度和测量不确定度均改善。
- 仍需更多努力确保中国临床实验室25(OH)D检测的可靠性和准确性。
研究设计与样本
- 研究制备7份不同25(OH)D浓度的血清池,发送至12家参与实验室。
- 各实验室先用现行LC-MS/MS法、后用优化LC-MS/MS法进行多次重复分析。
- 以变异系数评估精密度,以相对于NIST参考测量程序的偏倚评估正确度。
主要发现
- 现行方法下,25(OH)D2和25(OH)D3分别有43%和57%的实验室达到CV<12.5%和平均偏倚<8.3%的性能标准。
- 优化方法未确保所有实验室样本满足MU<13.6%标准,尤其低浓度样本,但较现行方法显著降低MU。
关键数值与推荐
- 性能标准:CV<12.5%,平均偏倚<8.3%。
- 测量不确定度标准:MU<13.6%。
临床怎么用
- 临床实验室可采用优化LC-MS/MS方法提升血清25(OH)D检测的精密度和正确度。
- 应关注低浓度样本的测量不确定度,必要时采取进一步改进措施。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Measuring serum 25-hydroxyvitamin D is key in clinical labs, but inter-laboratory variations risk diagnostic errors. This study evaluates the performance of current in-house liquid chromatography-tandem mass spectrometry (LC-MS/MS) methods used in top Chinese clinical laboratories and proposes an optimized method for improving serum 25(OH)D measurement accuracy and reliability. METHODS: Seven serum pools with different concentrations of 25(OH)D were prepared and sent to 12 participating laboratories for multiple repeat analysis with their current in-house LC-MS/MS methods and then an optimized LC-MS/MS method. Precision was assessed in terms of coefficient of variance (CV), and trueness was assessed in terms of bias referring to the U.S. National Institute of Standards and Technology (NIST) reference measurement procedure (RMP). The analytical performances of the two methods were compared and evaluated. RESULTS: Eighty percent and 90 % of the laboratories achieved the defined performance criteria (CV, <12.5 %; mean bias, <8.3 %) with the optimized method for the measurement of 25(OH)D2 and 25(OH)D3, compared with 43 % and 57 % of the laboratories meeting the criteria with their in-house methods, respectively. Precision and trueness improved after applying the optimized method. Although the optimized method didn't not ensure that all laboratory samples meet the measurement uncertainty (MU) criteria (MU<13.6 %), particularly for low-concentration samples, it significantly reduced the MU compared to the in-house method. CONCLUSIONS: Precision, trueness and MU improved after applying the optimized method. Nonetheless, more efforts are needed to ensure the reliability and accuracy of 25(OH)D measurements in clinical laboratories in China.
引用
- Zhou W, Gong M, Mao Y, et al. Analytical performance evaluation and optimization of serum 25(OH)D LC-MS/MS measurement. Clin Chem Lab Med. 2025. PMID: 39964255. doi:10.1515/cclm-2024-1416
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。