文献速递临床生化质量管理与危急值

按IFCC建议在AST和ALT检测中添加磷酸吡哆醛:填补数十年空白

Measurement of AST and ALT with Pyridoxal-5'-Phosphate according to IFCC: A decades-long gap seems to be filled.
PubMed 文献 · Clin Chim Acta · 2025-01-30

该研究比较了添加与不添加磷酸吡哆醛(PLP)测定转氨酶的方法,使用102份残留锂肝素样本在ILab Taurus仪器上检测。结果显示添加PLP的试剂与已含PLP的常规试剂结果高度一致,ALT平均升高12%;维生素B6较低患者中两种方法差值中位数显著更高(34 vs 18,p=0.006)。

要点

  • 添加PLP可提高转氨酶测定准确性,尤其维生素B6较低患者。
  • 添加PLP后ALT浓度平均升高12%,AST变化未报告。
  • 用户需适应含PLP试剂稳定性较低的问题并遵循国际学会建议。

核心结论

  • 添加PLP的试剂测定转氨酶准确性更高,尤其对维生素B6较低患者,可促进不同方法间结果可比性。

研究设计与样本

  • 收集102份已用含PLP试剂测定转氨酶的残留锂肝素样本,在ILab Taurus仪器上分别用含PLP(WP)和不含PLP(W)的Werfen试剂检测。
  • 维生素B6结果在相应残留K2-EDTA全血上获得,进行定性和定量统计分析。

主要发现

  • 含PLP的常规试剂(R)与WP之间具有最佳一致性和可比性。
  • 添加PLP使ALT浓度平均升高12%。
  • 维生素B6较低患者中WP-W差值中位数为34,较高患者为18(p=0.006)。

关键数值与推荐

  • ALT浓度平均升高12%(添加PLP vs 不添加)。
  • 维生素B6较低患者WP-W差值中位数34,较高患者18,p=0.006。

临床怎么用

  • 推荐在AST和ALT检测试剂中添加PLP作为辅因子,尤其对维生素B6较低患者。
  • 用户应调整操作流程以应对含PLP转氨酶试剂稳定性较低的问题,并遵循国际学会建议。

局限性

  • 样本为残留锂肝素血浆,维生素B6在K2-EDTA全血上检测,可能存在样本类型差异。
  • 仅使用Werfen试剂和ILab Taurus仪器,结果外推需谨慎。
  • 未报告AST的具体变化数值及样本量以外的统计细节。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Pyridoxal-5'-Phosphate (PLP) is not always added to the reagents for the measurement of Aspartate Amino Transferase (AST) and Alanine Amino Transferase (ALT), although the International Federation of Clinical Chemistry recommends using it as a cofactor. This is true for patients with lower Vitamin B6, precursor of PLP. The aim of this study is to compare the methods for transaminase measurement with or without the presence of PLP. 102 leftover LiHe samples were collected after transaminase measurement with the reagent already including PLP (R). The residual plasma was analyzed on the ILab Taurus instrumentation (Werfen, Spain) using Werfen reagents with (WP) and without (W) PLP. Vitamin B6 results were obtained on the respective residual K2-EDTA whole blood. Qualitative and quantitative statistical analyses were performed. Optimal agreement and comparability were found between R and WP. Addition of PLP to the reagents shows an average percentage increase in ALT concentrations of 12 %. Patients were categorized for Vitamin B6: the median value of WP-W for patients with lower Vitamin B6 is 34, 18 for patients with higher Vitamin B6 (p = 0.006), demonstrating that ALT values are higher when measured with WP for patients with low Vitamin B6. The addition of PLP to reagents has greater accuracy in transaminase determination, promoting optimal comparability of the results between different methods, particularly in patients' samples with reduced Vitamin B6. Users should adapt their procedures to overcome the difficulties related to the lower stability of transaminase reagents containing PLP, respecting also the recommendations of international scientific societies.

来源:Clin Chim Acta,查看原文。版权归原作者所有。

引用

  1. Talli I, Marchioro L, Zaninotto M, et al. Measurement of AST and ALT with Pyridoxal-5'-Phosphate according to IFCC: A decades-long gap seems to be filled. Clin Chim Acta. 2025. PMID: 39892693. doi:10.1016/j.cca.2025.120158

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。