真菌PCR倡议与Modimucor研究组开展的室间比对:提升血清中毛霉目DNA qPCR检测标准化
该研究向26家实验室分发模拟样本Panel A(6份加毛霉目DNA血清和1份阴性对照)和Panel B(6份毛霉目DNA提取物),评估DNA提取与PCR扩增参数对毛霉目qPCR效率的影响。Panel A阳性率在MucorGenius、Fungiplex和自建qPCR中分别为64%、70%和89%;使用1 mL大体积血清提取DNA的灵敏度为82.5%,高于小体积的62.7%(P<0.01)。Panel B中较大PCR输入体积(≥7 µL)的灵敏度为95.5%,高于低体积的58.3%(P<0.01)。
要点
- 使用1 mL大体积血清进行DNA提取可显著提高毛霉目qPCR分析灵敏度(82.5% vs 62.7%,P<0.01)。
- PCR扩增时使用≥7 µL较大DNA模板体积可显著提高灵敏度(95.5% vs 58.3%,P<0.01)。
- 提取平台和qPCR方法影响阳性率,但预分析变量(血清体积和模板体积)是灵敏度的主要决定因素。
核心结论
- 使用较大血清体积进行核酸提取和较大DNA模板体积进行PCR扩增,可显著改善血清毛霉目qPCR的检测性能。
研究设计与样本
- 真菌PCR倡议毛霉目实验室工作组向26家实验室分发两个模拟样本Panel:Panel A含6份加毛霉目DNA血清和1份阴性对照血清,Panel B含6份毛霉目DNA提取物。
- Panel A在各实验室按本地流程提取DNA后送至中心实验室,用三种qPCR技术检测:一种自建qPCR和两种商业试剂(MucorGenius和Fungiplex)。
- Panel B DNA提取物在各实验室按本地流程PCR扩增:九种自建qPCR和两种商业试剂盒(MucorGenius和MycoGENIE)。
- 所有数据在中心实验室匿名汇总分析。
主要发现
- Panel A使用六种不同自动化平台和五种手动提取方法。
- Panel A阳性率:MucorGenius为64%,Fungiplex为70%,自建qPCR为89%。
- Panel A中1 mL血清提取DNA的灵敏度为82.5%,小体积为62.7%(P<0.01)。
- Panel B中五种自建qPCR和两种商业试剂盒阳性率>78%。
- Panel B中PCR输入体积≥7 µL的灵敏度为95.5%,低输入体积为58.3%(P<0.01)。
临床怎么用
- 检测血清毛霉目游离DNA时,应优先采用大体积血清(如1 mL)进行DNA提取以提高分析灵敏度。
- PCR扩增时应使用较大DNA模板体积(≥7 µL)以提高检测灵敏度。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
UNLABELLED: The aim of this study was to identify parameters influencing DNA extraction and PCR amplification efficiencies in an attempt to standardize Mucorales qPCR. The Fungal PCR Initiative Mucorales Laboratory Working Group distributed two panels of simulated samples to 26 laboratories: Panel A (six sera spiked with Mucorales DNA and one negative control serum) and Panel B (six Mucorales DNA extracts). Panel A underwent DNA extraction in each laboratory according to the local procedure and were sent to a central laboratory for testing using three different qPCR techniques: one in-house qPCR assay and two commercial assays (MucorGenius and Fungiplex). Panel B DNA extracts were PCR amplified in each laboratory using local procedures: nine in-house qPCR assays and two commercial kits (MucorGenius and MycoGENIE). All data were compiled and anonymously analyzed at the central laboratory. For Panel A, a total of six different automated platforms and five manual extraction methods were used. Positive rates were 64%, 70%, and 89%, for the MucorGenius, Fungiplex, and the in-house qPCR assay, respectively. Using a large volume of serum for DNA extraction provided the highest analytical sensitivity (82.5% for 1 mL compared with 62.7% for smaller volumes, P < 0.01). For Panel B, five in-house qPCR assays and two commercial kits had >78% positivity. Using larger PCR input volumes (≥7 µL) was associated with the highest sensitivity at 95.5% compared to 58.3% when lower input volumes were used (P < 0.01). Using larger sample volumes for nucleic acid extraction and DNA template volumes for PCR amplification significantly improves the performance of Mucorales qPCR when testing serum. IMPORTANCE: Mucormycosis is a life-threatening mold infection affecting immunosuppressed patients but also other patients with diabetes or trauma. Better survival is linked to shorter delays in diagnosis and treatment initiation. Detection of Mucorales-free DNA in serum or plasma using quantitative PCR allows a prompt diagnosis and earlier treatment. Several techniques and protocols of quantitative Mucorales PCR are used in Europe, and improving performance remains a common objective of laboratories participating in the fungal PCR Initiative Working Group. This study, which combined results from 26 laboratories in Europe, showed that the main parameters underpinning sensitivity are the preanalytical variables (volume of serum used for DNA extraction and DNA template volume), irrespective of the extraction platforms and qPCR assay/platform.
引用
- Rocchi S, Scherer E, White PL, et al. Interlaboratory assays from the fungal PCR Initiative and the Modimucor Study Group to improve qPCR detection of Mucorales DNA in serum: one more step toward standardization. J Clin Microbiol. 2024. PMID: 39745482. doi:10.1128/jcm.01525-24