文献速递临床微生物分子诊断

Qvella FAST系统与常规方法在单微生物阳性血培养快速鉴定及药敏试验中的性能比较

Comparative performances of the Qvella FAST system and conventional methods for rapid identification and antibiotic susceptibility testing on monomicrobial positive blood cultures.
PubMed 文献 · J Clin Microbiol · 2024-12-20

本研究评估Qvella FAST系统在阳性血培养瓶中的快速鉴定与药敏性能,并与基于24小时传代培养的参考方法及常规快速手工流程比较。共纳入209份样本(107份前瞻性脓毒症患者血培养和102份掺杂难鉴定菌及多重耐药菌的血培养)。FAST系统种水平鉴定准确率为85.2%(178/209),药敏分类一致率99.4%,极重大误差0.59%;常规手工流程鉴定准确率86.2%、分类一致率99.6%,两者性能相似。

要点

  • FAST系统可在40分钟内从阳性血培养瓶制备液态菌落用于快速鉴定和药敏。
  • FAST系统种水平鉴定准确率85.2%,药敏分类一致率99.4%,极重大误差0.59%。
  • 常规快速手工流程性能与FAST系统相似,鉴定准确率86.2%、分类一致率99.6%。

核心结论

  • Qvella FAST系统在单微生物阳性血培养的快速鉴定和药敏试验中表现优异,可将诊断时间缩短约1天。

研究设计与样本

  • 研究比较FAST系统工作流程与常规快速手工流程,以基于24小时传代培养的参考方法为对照。

主要发现

  • FAST系统在178/209(85.2%)的病例中给出正确的种水平鉴定。
  • FAST系统药敏分类一致率为99.4%,极重大、重大和次要误差率分别为0.59%、0.20%和0.26%。
  • 常规快速手工流程鉴定准确率为86.2%,药敏分类一致率99.6%,极重大、重大和次要误差率分别为0.50%、0.16%和0.13%。

临床怎么用

  • FAST系统可用于单微生物阳性血培养的快速鉴定和药敏,有助于缩短脓毒症诊断时间。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Rapid and accurate diagnosis of sepsis is of paramount importance to reduce associated morbidity and mortality. The Qvella FAST System is a new instrument that concentrates and purifies bacteria from positive-flagged blood culture bottles (PFBCBs) to produce a "liquid" colony comparable to a subcultured colony in less than 40 min for rapid ID and calibrated antibiotic susceptibility testing (AST). In this study, we evaluated performances of the FAST System workflow and our rapid routine manual workflow (bacterial pellet obtained after lysis, cleaning, washing, and centrifugation for ID; AST by disc diffusion by direct inoculation after dilution) by comparison to the reference method based on 24-h bacterial subcultures. Two panels of PFBCBs were studied: panel A (including 107 prospective BCs from septic patients, October-November 2022) and panel B (including 102 BCs spiked with difficult-to-identify bacteria [mostly streptococci] and multidrug-resistant isolates), resulting in a total of 209 evaluable samples. The FAST System provided a correct ID to the species level in 178/209 (85.2%) of cases. For AST, the categorical agreement (CA) of the FAST System was 99.4%, with rates of very major (VME), major (ME), and minor (mE) errors of 0.59%, 0.20%, and 0.26%, respectively. Our rapid routine workflow based on manual methods show similar results for ID (86.2%) and AST (CA, 99.6%; VME, 0.50%; ME, 0.16%; mE, 0.13%). In conclusion, the Qvella FAST system, a promising tool that can reduce diagnostic time by approximately 1 day, shows excellent performances for rapid ID and AST.

来源:J Clin Microbiol,查看原文。版权归原作者所有。

引用

  1. Penven M, Louazon M, Freret C, et al. Comparative performances of the Qvella FAST system and conventional methods for rapid identification and antibiotic susceptibility testing on monomicrobial positive blood cultures. J Clin Microbiol. 2024. PMID: 39704520. doi:10.1128/jcm.01332-24

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。