15N标记重组蛋白同位素稀释UPLC-MS/MS候选参考方法绝对定量人血清胱抑素C
本研究以全长胱抑素C纯品和完全重组15N标记胱抑素C内标经蛋白酶切后,建立LC-MS/MS绝对定量血清胱抑素C的方法,溯源至氨基酸参考测量程序。方法总不精密度≤8.2%,回收率97.36–103.26%,对ERM-DA471-IFCC的相对偏倚1.74%,线性范围0.21–10.13 mg/L(R²=0.999)。
要点
- 建立了可溯源至氨基酸参考测量程序的血清胱抑素C候选参考方法。
- 方法总不精密度≤8.2%,对ERM-DA471-IFCC相对偏倚1.74%。
- 线性范围0.21–10.13 mg/L,R²=0.999,可用于评估免疫测定结果的准确性。
核心结论
- 建立了高选择性、准确的血清胱抑素C LC-MS/MS候选参考方法,可用于评估免疫测定结果与预期靶值的相关性。
研究设计与样本
- 以全长胱抑素C纯品和完全重组15N标记胱抑素C内标进行蛋白酶切,建立LC-MS/MS绝对定量方法。
- 使用氨基酸参考测量程序赋值的校准品进行溯源,并以已知氨基酸浓度的氨基酸参考物质校准仪器进行比较。
主要发现
- 方法总不精密度≤8.2%。
- 回收率范围为97.36–103.26%。
- 对ERM-DA471-IFCC的候选参考测量程序测定值与靶值相对偏倚为1.74%。
- 线性响应范围为0.21–10.13 mg/L,R²为0.999。
- 本方法结果与其他认证参考测量程序结果一致。
关键数值与推荐
- 总不精密度≤8.2%。
- 回收率97.36–103.26%。
- 对ERM-DA471-IFCC相对偏倚1.74%。
- 线性范围0.21–10.13 mg/L,R²=0.999。
临床怎么用
- 当临床怀疑血清胱抑素C免疫测定结果存在差异时,可用该方法评估免疫测定结果与预期靶值的相关性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Serum cystatin C (CysC) is a reliable and ideal endogenous marker for accurately assessing early changes in glomerular filtration rate (GFR), surpassing the limitations of creatinine-based estimated GFR. To improve the precision of GFR calculation, the development of strategies for accurately measuring serum CysC is crucial. METHODS: In this study, the full-length CysC pure product and fully recombinant 15N-labeled CysC internal standard were subjected to protein cleavage. Subsequently, an LC-MS/MS method was developed for the absolute quantification of serum CysC. The traceability of the method was assigned calibrator using the amino acid reference measurement procedure (RMP). It involved calibrating the instrument using an amino acid reference material with known amino acid concentrations for calibration and comparison purposes. RESULTS: The total imprecision of the method was determined to be ≤8.2 %, and a lower functional limit of quantification (LLoQ) was achieved. The recoveries ranged from 97.36 to 103.26 %. The relative bias between this candidate RMP for measurement of ERM-DA471-IFCC and the target value was 1.74 %. The linearity response was observed within the concentration range of 0.21-10.13 mg/L, with a high R2 value of 0.999. The results obtained using our method was consistent with those obtained using other certified RMPs. CONCLUSIONS: With the establishment of this highly selective and accurate serum CysC measurement method, it is now possible to assess the correlation between immunoassay results of serum CysC and the intended target when discrepancies are suspected in the clinical setting.
引用
- Zhang Q, Zhan M, Peng X, et al. Absolute quantitation of human serum cystatin C: candidate reference method by 15N-labeled recombinant protein isotope dilution UPLC-MS/MS. Clin Chem Lab Med. 2024. PMID: 39323320. doi:10.1515/cclm-2024-0300