两种气动管道运输系统对血样溶血影响的比较
本研究比较了Tempus600 hub两步式单管系统与Airco一步式容器气动管道运输系统在院内运输血样时对溶血的影响。健康志愿者血样分装于10 mL采血管(全容量或半容量)经两种系统运输约250米,检测LDH、K+和溶血指数。Tempus600 hub系统导致溶血指数升高,仅LDH差异有统计学意义。
要点
- Tempus600 hub系统运输后溶血指数较Airco系统升高。
- 全容量和半容量样本溶血指数分别升高49%和46%。
- 常规样本中Tempus600 hub系统溶血指数中位数升高23%和33%。
核心结论
- Tempus600 hub系统导致基于溶血指数的不可报告LDH或K+结果增多。
- 仅LDH差异有统计学意义(p<0.01和p<0.08),K+差异无统计学意义(p<0.28和p<0.56)。
研究设计与样本
- 来自健康志愿者的10份血样分装于10 mL采血管,全容量或半容量填充,经两种气动管道运输系统运输约250米。
- 检测溶血标志物LDH、K+和溶血指数,并调查常规样本和重复运输中溶血指数的差异。
- 使用冲击数据记录仪记录两种系统的加速度曲线以评估机械冲击影响。
主要发现
- Tempus600 hub系统运输后,全容量和半容量样本的溶血指数分别升高49%和46%。
- 常规样本经Tempus600 hub系统运输后溶血指数中位数升高23%和33%。
- 冲击记录仪分析显示Tempus600 hub系统冲击次数增加6.5倍,冲击强度增加1.8倍。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Efficient and timely transportation of clinical samples is pivotal to ensure accurate diagnoses and effective patient care. During the transportation process, preservation of sample integrity is crucial to avoid pre-analytical aberrations on laboratory results. Here, we present a comparative analysis between a two-step Tempus600 hub solution single-tube and a one-step, container-based pneumatic transport system (PTS) from Airco, for the in-house transportation of blood samples. METHODS: Ten blood samples from healthy volunteers were split in 10 mL collection tubes filled at full or half capacity for transportation with the two PTS (about 250 m). To compare the impact of transportation, markers of hemolysis such as lactate dehydrogenase (LDH), potassium (K+), and the hemolysis index (HI), were determined. Additionally, differences in HI in routine samples and repeated transportation was investigated. To assess and compare the mechanistic impact profiles, we recorded the acceleration profiles of the two PTS using a shock data logger. RESULTS: Transportation using the Tempus600 hub solution resulted in 49 and 46 % higher HI with samples filled to total or half capacity, respectively. Routine samples transported with the Tempus600 hub solution showed a higher median HI by 23 and 33 %. Additionally, shock logger analysis showed an elevated amount of shocks (6.5 fold) and shock intensities (1.8 fold). CONCLUSIONS: The Tempus600 hub solution caused an increased number of unreportable LDH or K+ results based on the hemolysis index. However, it was only statistically significant for LDH (p<0.01 and p<0.08) - while the comparisons for K+ were not statistically significant (p<0.28 and p<0.56).
引用
- Luginbühl M, Frey K, Gawinecka J, et al. Comparison of a two-step Tempus600 hub solution single-tube vs. container-based, one-step pneumatic transport system. Clin Chem Lab Med. 2024. PMID: 38742247. doi:10.1515/cclm-2024-0057