文献速递临床微生物质量管理与危急值

Selux下一代表型药敏系统的多中心评估

Multicenter evaluation of the Selux Next-Generation Phenotyping antimicrobial susceptibility testing system.
PubMed 文献 · J Clin Microbiol · 2023-12-05

Selux NGP系统是一种新型药敏试验系统,通过双序列检测确定MIC。多中心评估按FDA建议并与参考肉汤微量稀释法比较,纳入2488株临床和挑战菌株,革兰阴性菌测24种抗菌药,革兰阳性菌测15种。总体极重大错误和重大错误率均低于1%,必要一致性和分类一致性均>95%,重现性≥95%,平均出结果时间5.65小时。

要点

  • 总体极重大错误率<1%(31/3805),重大错误率<1%(107/15606)。
  • 必要一致性和分类一致性均>95%,重现性≥95%。
  • 平均出结果时间(从试验开始到MIC结果)为5.65小时。

核心结论

  • Selux NGP系统与参考肉汤微量稀释法比较,总体极重大错误和重大错误率均低于1%,必要一致性和分类一致性均>95%,重现性≥95%,平均出结果时间5.65小时。

研究设计与样本

  • 多中心评估,遵循FDA建议并使用FDA定义的折点,将Selux NGP系统与参考肉汤微量稀释法比较。
  • 共纳入2488株临床和挑战分离株;革兰阴性分离株测试24种抗菌药,革兰阳性分离株测试15种抗菌药。

主要发现

  • 总体极重大错误率为31/3805(<1%),重大错误率为107/15606(<1%)。
  • 必要一致性和分类一致性均>95%,重现性≥95%。
  • 平均出结果时间(从试验开始到MIC结果)为5.65小时。
  • 数据涵盖了所有评估的菌种-抗菌药组合,包括符合和不符合FDA性能要求的组合。

关键数值与推荐

  • 极重大错误率<1%(31/3805),重大错误率<1%(107/15606)。
  • 必要一致性>95%,分类一致性>95%,重现性≥95%。
  • 平均出结果时间5.65小时。

临床怎么用

  • 该系统可快速(平均5.65小时)提供MIC结果,且与参考方法一致性高,可能有助于临床及时调整抗感染治疗。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

The Selux Next-Generation Phenotyping (NGP) system (Charlestown, MA) is a new antimicrobial susceptibility testing system that utilizes two sequential assays performed on all wells of doubling dilution series to determine MICs. A multicenter evaluation of the performance of the Selux NGP system compared with reference broth microdilution was conducted following FDA recommendations and using FDA-defined breakpoints. A total of 2,488 clinical and challenge isolates were included; gram-negative isolates were tested against 24 antimicrobials, and gram-positive isolates were tested against 15 antimicrobials. Data is provided for all organism-antimicrobial combinations evaluated, including those that did and did not meet FDA performance requirements. Overall very major error and major error rates were less than 1% (31/3,805 and 107/15,606, respectively), essential agreement and categorical agreement were >95%, reproducibility was ≥95%, and the average time-to-result (from time of assay start to time of MIC result) was 5.65 hours.

来源:J Clin Microbiol,查看原文。版权归原作者所有。

引用

  1. Baker KR, Flentie K, Spears BR, et al. Multicenter evaluation of the Selux Next-Generation Phenotyping antimicrobial susceptibility testing system. J Clin Microbiol. 2023. PMID: 38051069. doi:10.1128/jcm.00546-23

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。