全血HCV RNA逆转录环介导等温扩增快速检测方法的建立
研究开发并评估三种从少量全血中提取HCV RNA的简易流程,结合泛基因型RT-LAMP检测。水流程、热流程和磁珠流程的检测限分别为4.3、4.2和2.8 log10(IU/mL),对低病毒载量样本的检出率分别为69%、75%和94%,每个流程可在60分钟内完成且兼容侧向层析试纸条。
要点
- 磁珠流程检测限最低(2.8 log10 IU/mL),低病毒载量样本检出率94%
- 三种流程均可在60分钟内完成,设备需求少且成本低
- 流程兼容可视化侧向层析试纸条,有望用于即时诊断
核心结论
- 三种全血HCV RNA提取流程结合RT-LAMP检测简单、快速、灵敏,磁珠流程性能最优。
研究设计与样本
- 开发并评估三种从少量全血中提取HCV RNA的流程(水、热、磁珠),用于泛基因型RT-LAMP检测。
主要发现
- 水流程检测限为4.3 log10(IU/mL)(95% CI 4.0-4.9)。
- 热流程检测限为4.2 log10(IU/mL)(95% CI 3.8-5.1)。
- 磁珠流程检测限为2.8 log10(IU/mL)(95% CI 2.5-3.4)。
- 对多数HCV水平低于5.0 log10(IU/mL)的掺入全血样本,水、热、磁珠流程检出率分别为69%、75%、94%。
临床怎么用
- 三种流程设备需求少、成本低,可在60分钟内出结果,兼容侧向层析试纸条,适合非实验室环境使用。
局限性
- 需进一步简化操作流程,方可作为HCV感染即时诊断检测的基础。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Hepatitis C virus (HCV) infection is an underdiagnosed global health problem. Diagnosis of current HCV infections typically requires testing for HCV RNA using high-complexity laboratory tests. Methods for the detection of HCV RNA that are simple, inexpensive, rapid, and compatible with use outside of a laboratory setting are very important in order to improve access to hepatitis C diagnostic testing and facilitate accelerated linkage to care. We developed and evaluated three simple workflows for extracting HCV RNA from small volumes of whole blood for use in a sensitive, pan-genotypic RT-LAMP assay. The water workflow uses osmotic stress to release HCV RNA and has a limit of detection of 4.3 log10(IU/mL) (95% CI 4.0-4.9). The heat workflow uses a heating step to release HCV RNA and has a limit of detection of 4.2 log10(IU/mL) (95% CI 3.8-5.1). The bead workflow, which uses chemical lysis of the sample and a streamlined paramagnetic solid phase reversible immobilization bead procedure for nucleic acid purification, has a limit of detection of 2.8 log10(IU/mL) (95% CI 2.5-3.4). When used to test whole blood spiked with HCV RNA-positive plasma samples in which most HCV levels were below 5.0 log10(IU/mL), the water, heat, and bead workflows detected HCV RNA in 69%, 75%, and 94% of samples, respectively. These workflows are compatible with visual lateral flow dipsticks, and each takes less than 60 min from sample to result. Each workflow can be performed with minimal and inexpensive equipment. With further procedural simplifications, these workflows may form the basis of assays for the point-of-care diagnosis of HCV infections.
引用
- Pauly MD, Weis-Torres S, Hayden TM, et al. Development of simple, rapid, and sensitive methods for detection of hepatitis C virus RNA from whole blood using reverse transcription loop-mediated isothermal amplification. J Clin Microbiol. 2023. PMID: 37933990. doi:10.1128/jcm.00771-23