血浆肾素活性测定候选参考物质互换性评估的现状与挑战

Commutability assessment of candidate reference materials for plasma renin activity measurement: current challenges.
PubMed 文献 · Clin Chem Lab Med · 2023-07-21

本研究评估了16种候选参考物质(RMs)和40份临床样本在四种常规血浆肾素活性(PRA)测定中的互换性,包括三种LC-MS/MS法和一种化学发光免疫分析法。采用偏差差异法和线性回归预测区间法判定互换性,并分析添加和混合规模的影响。结果显示互换性结果不理想(偏差差异法非互换性14/16),大规模混合池趋于可互换,当前EQA材料具有较好互换性、稳定性和均匀性,适合作为RMs。

要点

  • 当前EQA材料互换性较好且稳定均匀,适合作为PRA测定的参考物质。
  • 大规模混合池(>1000人)无论是否添加均趋于可互换,小规模添加池不建议用于制备RMs。
  • 采用统一孵育缓冲液的方法可能更有利于互换性研究。

核心结论

  • 可互换的参考物质可获得但具有挑战性,当前EQA材料适合作为RMs。

研究设计与样本

  • 研究评估了16种候选RMs和40份临床样本,使用四种常规PRA测定(三种LC-MS/MS和一种化学发光免疫分析)。
  • 候选RMs包括小规模混合池(<50人)和当前EQA材料(大规模混合池>1000人),采用偏差差异法和线性回归预测区间法判定互换性。

主要发现

  • 所有测定方法的精密度和相关性表现可接受。
  • 偏差差异法中互换性结果不理想(非互换性14/16),使用不同孵育缓冲液的测定对间检测到显著样本特异性效应。
  • 预测区间法中,添加的小规模混合池未观察到互换性;EQA材料(4/9)比小规模混合池(2/7)在各测定间有更满意的互换性。
  • 大规模混合池的RMs无论是否添加均趋于更好的互换性。

关键数值与推荐

  • 偏差差异法判定非互换性为14/16。
  • 预测区间法中EQA材料互换性满意比例为4/9,小规模混合池为2/7。

临床怎么用

  • 当前具有较好互换性、稳定性和均匀性的EQA材料可作为PRA测定的参考物质。
  • 不建议通过在小规模混合池中添加物质来制备参考物质。
  • 为进一步的互换性研究,建议方法采用统一的孵育缓冲液。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: This study aims to evaluate the commutability of external quality assessment (EQA) materials and candidate reference materials (RMs) for plasma renin activity (PRA) assay. METHODS: Commutabilities of 16 candidate RMs were measured along with 40 clinical samples by the four different routine PRA assays, including three LC‒MS/MS assays and one chemiluminescence immunoassay. Sixteen candidate RMs included native/spiked human plasma pools (small-scale pools with <50 individuals) and current EQA materials (large-scale pools with >1,000 individuals). Difference in bias approach and linear regression with prediction interval approach were adopted to determine the commutability. Two-way variance analysis was used to estimate the effects of spiked and pool size on the commutability. Stability and homogeneity studies were performed. RESULTS: Precision and correlation performance of all assays was acceptable. In the difference in bias approach, the commutability results were not satisfactory (noncommutability: 14/16) and significant sample-specific effects were detected in assay pairs using different incubation buffers. For the prediction interval approach, no commutability was observed in the spiked small-scale pools; EQA materials (4/9) had more satisfactory commutability among all assays than the small-scale pools (2/7); RMs of large-scale pools tend to have better commutability no matter spiked or not. CONCLUSIONS: Commutable RMs were obtainable but challenging. Current EQA materials with relatively good commutability, stability, and homogeneity were appropriate RMs. Large-scale pools are tending to be commutable. Spiking in small-scale pools was not suggested to prepare RMs. MPs adopting a uniform incubation buffer would be preferable for further commutability research.