舒洛培南纸片开发、质控范围及肠杆菌目分离株MIC与纸片结果相关性
研究评估了2、5、10和20 μg舒洛培南纸片的药敏试验,选用2 μg纸片,依据CLSI M23指南以≤0.5/1/≥2 μg/mL的S/I/R判读标准进行误差率分析。在2856株肠杆菌目中极少判读误差(无极重大误差,仅1个重大误差)。八实验室质控研究显示99.0%(470/475)结果在24-30 mm的7 mm范围内,CLSI建立了大肠埃希菌29522的2 μg纸片质控范围24-30 mm。
要点
- 舒洛培南2 μg纸片对肠杆菌目药敏试验准确且重复性好。
- 采用≤0.5/1/≥2 μg/mL判读标准时极少数判读误差。
- 大肠埃希菌29522的2 μg纸片质控范围为24-30 mm。
核心结论
- 舒洛培南2 μg纸片用于肠杆菌目药敏试验准确且可重复。
研究设计与样本
- 评估2、5、10和20 μg舒洛培南纸片,选用2 μg纸片,对2856株肠杆菌目进行肉汤微量稀释法和纸片扩散法药敏试验,依据CLSI M23指南进行误差率分析。
- 八实验室质控研究使用2 μg纸片,评估质控结果范围。
主要发现
- 在2856株肠杆菌目中,无极重大误差,仅1个重大误差。
- 八实验室质控研究中99.0%(470/475)结果在24-30 mm的7 mm范围内,不同纸片批号和培养基结果相似,无异常站点。
关键数值与推荐
- 提出的舒洛培南敏感/中介/耐药判读标准为≤0.5/1/≥2 μg/mL。
- CLSI建立的大肠埃希菌29522对2 μg舒洛培南纸片的质控范围为24-30 mm。
临床怎么用
- 临床微生物实验室可使用2 μg舒洛培南纸片按CLSI标准对肠杆菌目进行药敏试验,并以大肠埃希菌29522进行质控。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Sulopenem disk masses of 2, 5, 10, and 20 μg were evaluated by susceptibility testing isolates by broth microdilution and disk diffusion. A 2-μg disk was chosen, and error-rate bounding analysis in accordance with Clinical and Laboratory Standards Institute (CLSI) guideline M23 was conducted using a proposed sulopenem susceptible/intermediate/resistant (S/I/R) interpretive criterion of ≤0.5/1/≥2 μg/mL. Among the evaluated Enterobacterales (n = 2,856), very few interpretive errors were observed (no very major errors and only one major error). An eight-laboratory quality control (QC) study was performed using the 2-μg disk, and 99.0% (470/475) of results were within a 7-mm range of 24 to 30 mm. Results were similar by disk lot and media, and no outlier sites were observed. A sulopenem 2-μg disk QC range for Escherichia coli 29522 of 24 to 30 mm was established by the CLSI. A 2-μg sulopenem disk performs accurately and reproducibly for testing of Enterobacterales.
引用
- Puttagunta S, Aronin SI, DiFranco-Fisher J, et al. Sulopenem Disk Development, Quality Control Range, and MIC to Disk Result Correlation for Enterobacterales Isolates. J Clin Microbiol. 2023. PMID: 37358462. doi:10.1128/jcm.00246-23
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。