两种胱抑素C试剂与四种方程对标准化后肾小球滤过率估算的影响
该研究基于148名参与者,模拟分析两种标准化胱抑素C试剂与四种eGFR方程组合的效果。结果显示两试剂在低浓度(<0.85 mg/L)等效性不佳,不同测量系统与方程组合可导致eGFR差异增大。
要点
- 两试剂在cysC低于0.85 mg/L时等效性不令人满意。
- 不同测量系统与方程组合可导致eGFR差异增大。
- eGFR的Lin一致性相关系数随组合不同在0.73至1.00间变化。
核心结论
- 两种胱抑素C试剂在低浓度(<0.85 mg/L)的等效性不令人满意,不同测量系统与方程组合可导致eGFR差异增大。
研究设计与样本
- 模拟分析,纳入148名参与者(平均年龄60.5±14.5岁,43%女性)。
- 使用Gentiancys和Rochecys两种试剂在Cobas c702系统上测量cysC。
- eGFR由四种方程(CKD-EPIcys、CAPAeq、FASeq、EKFCcys)的八种组合生成。
主要发现
- 平均cysC:Gentiancys为1.72±1.44 mg/L,Rochecys为1.71±1.35 mg/L。
- 回归分析显示在±7.61%总允许误差下,两试剂在0.85-4.40 mg/L范围内一致。
- eGFR的Lin一致性相关系数随测量系统与方程组合不同,范围为0.73至1.00。
临床怎么用
- 解读eGFR时需考虑cysC试剂与估算方程的具体组合,不同组合结果可能存在差异。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Serum cystatin C (cysC), which is less affected by sex, race, and muscle mass than creatinine, is a useful biomarker of the estimated glomerular filtration rate (eGFR). The standardization of cysC measurements remains controversial, although a certified reference material (ERM-DA471/IFCC) is available. Moreover, the effect of combinations of cysC reagents and equations for eGFR is unclear. METHODS: We conducted a simulation analysis of cysC measured using two reagents standardized against ERM-DA471/IFCC-Gentian cystatin C immunoassay (Gentiancys; GentianAS, Moss, Norway) and Roche Tina-quant Cystatin C Gen.2 (Rochecys; Roche, Mannheim, Germany)-on a Cobas c702 system (Roche) and eGFR generated by eight combinations of four equations: 2012 cystatin C-based Chronic Kidney Disease Epidemiology Collaboration equation (CKD-EPIcys); the Caucasian, Asian, pediatric, and adult equation (CAPAeq); full age spectrum equation (FASeq); and 2023 cystatin C-based European Kidney Function Consortium equation (EKFCcys). RESULTS: A total of 148 participants (mean age, 60.5±14.5 years; 43% female) were enrolled. The mean cysC was 1.72±1.44 mg/L for Gentiancys and 1.71±1.35 mg/L for Rochecys. Regression analysis showed concordance between the reagents within 0.85-4.40 mg/L when using ±7.61% total allowable error. Lin's concordance correlation coefficient of eGFR, by combining the measuring system and equation, varied from 0.73 to 1.00. CONCLUSIONS: The equivalence of cysC values at low concentrations (<0.85 mg/L) between the two reagents was unsatisfactory. Results obtained with different measurement systems could lead to larger differences in eGFR varying with the combination.
引用
- Lee HS, Bae GE, Lee JE, et al. Effect of Two Cystatin C Reagents and Four Equations on Glomerular Filtration Rate Estimations After Standardization. Ann Lab Med. 2023. PMID: 37387489. doi:10.3343/alm.2023.43.6.565
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。