液相色谱质谱法评估睾酮、雌二醇和孕酮免疫分析校准品

Evaluation of testosterone, estradiol and progesterone immunoassay calibrators by liquid chromatography mass spectrometry.
PubMed 文献 · Clin Chem Lab Med · 2023-06-05

该研究采用超高压液相色谱-质谱联用(UPLC-MS)测定四家制造商提供的41份睾酮(T)、雌二醇(E2)和孕酮(P4)免疫分析校准品的甾体浓度。结果显示部分校准品测定值与标签浓度显著偏差,提示校准不准确可能影响直接免疫分析准确性。

要点

  • 睾酮校准品14份中6份(43%)显著偏离标签浓度。
  • 雌二醇校准品14份中8份(57%)显著偏离标签浓度。
  • 孕酮校准品11份中8份(73%)显著偏离标签浓度。

核心结论

  • 制造商提供的校准品校准不准确可能导致商业直接甾体免疫分析不准确。

研究设计与样本

  • 采用UPLC-MS测定四家制造商提供的41份睾酮、雌二醇和孕酮免疫分析校准品(每份4-6个重复)的甾体浓度。

主要发现

  • 14份非零睾酮校准品中6份(43%)显著偏离标签浓度,其中29%超出标签浓度20%范围。
  • 14份雌二醇校准品中8份(57%)显著偏离,其中7份(50%)超出标签浓度20%范围。
  • 11份孕酮校准品中8份(73%)显著偏离,其中4份(36%)在标签浓度20%范围内。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: In clinical practice, steroid measurements are performed mainly by direct, non-extraction immunoassays adapted to high throughput, automated immunoassay platforms and employing secondary calibrators. The accuracy of such steroid immunoassays is limited by cross-reactivity with structurally related steroids and nonspecific matrix interference as well as the metrological traceability of manufacturer supplied calibrators. The accuracy of steroid immunoassay calibrators has been little investigated by independent chemical methods. METHODS: Steroid concentrations of 41 calibrators (4-6 replicates per calibrator) supplied by four manufacturers for use in testosterone (T), estradiol (E2), and progesterone (P4) commercial immunoassays were measured by ultra-pressure liquid chromatography-mass spectrometry (UPLC-MS). RESULTS: Among 14 non-zero T calibrators, six (43 %) deviated significantly from the label concentration with 29 % outside 20 % of it. Among 14 E2 calibrators, eight (57 %) deviated significantly, whereas seven (50 %) were outside 20 % of the label concentration. Among 11 P4 calibrators, eight (73 %) deviated significantly whereas four (36 %) were outside within 20 % of the label concentration. CONCLUSIONS: We conclude that inaccurate calibration of manufacturer's supplied standards may contribute to inaccuracy of commercial direct steroid immunoassays.

来源:Clin Chem Lab Med,查看原文。版权归原作者所有。