文献速递临床免疫质量管理与危急值

溶血、脂血和黄疸对常见临床免疫测定的干扰评估

Evaluation of hemolysis, lipemia, and icterus interference with common clinical immunoassays.
PubMed 文献 · Clin Chem Lab Med · 2023-01-13

研究使用残余患者血清池添加已知干扰物,在Alinity i分析仪上评估超过25项免疫测定的HIL干扰阈值。结果显示C肽、雌二醇、叶酸、游离T4、同型半胱氨酸、胰岛素和维生素B12受溶血影响,BNP、雌二醇、游离T3和同型半胱氨酸受黄疸影响,BNP、叶酸和同型半胱氨酸受脂血影响,HIV和肝炎血清学测定也受干扰。

要点

  • 溶血在血红蛋白0.3-20 g/L影响C肽、雌二醇、叶酸、游离T4、同型半胱氨酸、胰岛素和维生素B12。
  • 黄疸在结合胆红素50-1,044 μmol/L影响BNP、雌二醇、游离T3和同型半胱氨酸。
  • 脂血在甘油三酯0.8-10 mmol/L影响BNP、叶酸和同型半胱氨酸。

核心结论

  • 免疫测定受不同程度HIL干扰,部分项目受影响方式此前未报道。

研究设计与样本

  • 使用残余患者血清池添加溶血物、胆红素和Intralipid模拟内源性HIL,在Alinity i分析仪上分析超过25项免疫测定。

主要发现

  • 溶血在血红蛋白0.3-20 g/L影响C肽、雌二醇、血清叶酸、游离T4、同型半胱氨酸、胰岛素和维生素B12。
  • 黄疸在结合胆红素50-1,044 μmol/L影响BNP、雌二醇、游离T3和同型半胱氨酸。
  • 脂血在甘油三酯0.8-10 mmol/L影响BNP、血清叶酸和同型半胱氨酸。
  • HIV和甲、乙、丙型肝炎血清学免疫测定也受干扰。

临床怎么用

  • 为资源有限、无法内部验证内源性干扰的实验室提供Alinity i分析仪的HIL阈值参考。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Hemolysis, icterus, and lipemia (HIL) are common sources of endogenous interference in clinical laboratory testing. Defining the threshold of interference for immunoassays enables appropriate reporting of their results when they are affected by HIL. METHODS: Pools of residual patient serum samples were spiked with a known amount of interferent to create samples with varying concentrations of hemolysate, bilirubin, and Intralipid that mimicked the effects of endogenous HIL. Samples were analysed on the Alinity i analyser (Abbott Diagnostics) for more than 25 immunoassays. The average recovery relative to the non-spiked sample was calculated for each interference level and was compared to a predefined allowable bias. RESULTS: C-peptide, estradiol, serum folate, free T4, homocysteine, insulin, and vitamin B12 were found to be affected by hemolysis, at hemoglobin concentrations between 0.3 to 20 g/L. Immunoassays for BNP, estradiol, free T3, and homocysteine were affected by icterus at conjugated bilirubin concentrations between 50 to 1,044 μmol/L. BNP, serum folate, and homocysteine were affected by Intralipid with measured triglyceride concentrations between 0.8 to 10 mmol/L. Lastly, serological immunoassays for HIV and hepatitis A, B and C were also affected by interferences. CONCLUSIONS: Immunoassays are impacted by varying degrees of HIL interference. Some measurands, in the presence of interference, are affected in a manner not previously indicated. The data presented herein provide an independent evaluation of HIL thresholds and will be of aid to resource-limited clinical laboratories that are unable to internally verify endogenous interferences when implementing the Alinity i analyser.

来源:Clin Chem Lab Med,查看原文。版权归原作者所有。

引用

  1. Karin A, Higgins V, Miller J, et al. Evaluation of hemolysis, lipemia, and icterus interference with common clinical immunoassays. Clin Chem Lab Med. 2023. PMID: 36635939. doi:10.1515/cclm-2022-0924

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。