基于质谱参考测量程序与免疫测定比较评估脂蛋白(a)候选参考物质的互换性
该研究比较了基于质谱的候选参考测量程序与免疫测定法在39份临床样本中的相关性,并评估了14种候选参考物质的互换性。结果显示免疫测定法与候选参考测量程序对临床样本呈良好线性相关,但存在样本特异性差异;未加标人血清混合池可与临床样本互换,而加标重组载脂蛋白(a)的混合池行为不同。未加标人血清混合池是未来SI溯源脂蛋白(a)参考测量系统的首选候选次级参考物质。
要点
- 未加标人血清混合池是未来SI溯源Lp(a)参考测量系统的首选次级参考物质。
- 免疫测定法与候选参考测量程序对临床样本呈良好线性相关,但存在样本特异性差异。
- 加标重组载脂蛋白(a)的人血清混合池与临床样本的互换性行为不同。
核心结论
- 未加标人血清混合池是未来SI溯源脂蛋白(a)参考测量系统的首选候选次级参考物质。
研究设计与样本
- 研究评估了候选参考测量程序与免疫测定法在39份临床样本中的相关性,并调查了14种候选参考物质的互换性。
主要发现
- 免疫测定法与候选参考测量程序对临床样本呈良好线性相关,但也观察到一些显著的样本特异性差异。
- 基于未加标人血清混合池的参考物质可与临床样本互换。
- 加标重组载脂蛋白(a)的人血清混合池与临床样本相比表现出不同行为。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Elevated concentrations of lipoprotein(a) [Lp(a)] are directly related to an increased risk of cardiovascular diseases, making it a relevant biomarker for clinical risk assessment. However, the lack of global standardization of current Lp(a) measurement procedures (MPs) leads to inconsistent patient care. The International Federation for Clinical Chemistry and Laboratory Medicine working group on quantitating apolipoproteins by mass spectrometry (MS) aims to develop a next-generation SI (International system of units)-traceable reference measurement system consisting of a MS-based, peptide-calibrated reference measurement procedure (RMP) and secondary serum-based reference materials (RMs) certified for their apolipoprotein(a) [apo(a)] content. To reach measurement standardization through this new measurement system, 2 essential requirements need to be fulfilled: a sufficient correlation among the MPs and appropriate commutability of future serum-based RMs. METHODS: The correlation among the candidate RMP (cRMP) and immunoassay-based MPs was assessed by measuring a panel of 39 clinical samples (CS). In addition, the commutability of 14 different candidate RMs was investigated. RESULTS: Results of the immunoassay-based MPs and the cRMPs demonstrated good linear correlations for the CS but some significant sample-specific differences were also observed. The results of the commutability study show that RMs based on unspiked human serum pools can be commutable with CS, whereas human pools spiked with recombinant apo(a) show different behavior compared to CS. CONCLUSIONS: The results of this study show that unspiked human serum pools are the preferred candidate secondary RMs in the future SI-traceable Lp(a) Reference Measurement System.
引用
- Dikaios I, Althaus H, Angles-Cano E, et al. Commutability Assessment of Candidate Reference Materials for Lipoprotein(a) by Comparison of a MS-based Candidate Reference Measurement Procedure with Immunoassays. Clin Chem. 2023. PMID: 36644921. doi:10.1093/clinchem/hvac203
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。