校准频率与实验室检测的分析变异性
本研究通过模拟5点校准曲线(浓度10–10000 mmol/l,信号CV 3%)并设置3个QC水平(20、150、600 mmol/l),考察校准频率对检测变异性的影响。结果显示不同校准频率下QC的CV保持稳定;普通最小二乘回归时低浓度20 mmol/l不精密度最大(18%),而加权最小二乘回归时3个QC浓度CV均恒定在3.4%接近预设3%。
要点
- 增加校准频率不恶化整体分析性能。
- 低浓度(20 mmol/l)用普通最小二乘回归时不精密度达18%。
- 加权最小二乘回归使各QC浓度CV稳定在3.4%。
核心结论
- 在本模拟研究参数范围内,增加校准事件频率在多数情况下不会对测量程序整体分析性能产生不利影响。
研究设计与样本
- 模拟研究:构建5点校准曲线,模拟浓度10至10000 mmol/l,信号强度均值周围CV为3%且呈高斯分布。
- 模拟3个QC水平(20、150、600 mmol/l)穿插于分析测量范围内。
主要发现
- 3个QC水平的CV在不同校准频率(每校准周期5、10、15、30次QC测量)下保持稳定。
- 普通最小二乘回归时,20 mmol/l处不精密度最大为18%,150和600 mmol/l处分别降至3.5%和3.8%。
- 加权最小二乘回归时,3个QC浓度CV均恒定在3.4%,接近预设CV(3%)。
- 采用WLS模型的2点校准在较窄浓度范围(50和100 mmol/l以及50和500 mmol/l)观察到相似结果。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: There is uncertainty whether increased frequency of calibrations may affect the overall analytical variability of a measurement procedure as reflected in quality control (QC) performance. In this simulation study, we examined the impact of calibration frequencies on the variability of laboratory measurements. METHODS: A 5-point calibration curve was modeled with simulated concentrations ranging from 10 to 10,000 mmol/l, and signal intensities with CVs of 3 % around the mean, under a Gaussian distribution. 3 levels of QC (20, 150, 600 mmol/l) interspersed within the analytical measurement range were also simulated. RESULTS: The CV of the 3 QC levels remained stable across the different calibration frequencies simulated (5, 10, 15 and 30 QC measurements per recalibration episode). The imprecision was greatest (18 %) at the lowest concentration of 20 mmol/l, when the calibration curve was derived using ordinary least squares regression, reducing to 3.5 % and 3.8 % at 150 and 600 mmol/l, respectively. The CV of all 3 QC concentrations remained constant at 3.4 % and close the predefined CV (3 %) when weighted least squares regression was used to derive the calibration model. Similar findings were observed with 2-point calibrations using WLS models at narrower concentration ranges (50 and 100 mmol/l as well as 50 and 500 mmol/l). DISCUSSION: Within the parameters of the simulation study, an increased frequency of calibration events does not adversely impact the overall analytical performance of a measurement procedure under most circumstances.
引用
- Lim CY, Ow Yang S, Markus C, et al. Calibration frequency and analytical variability of laboratory measurements. Clin Chim Acta. 2022. PMID: 36513171. doi:10.1016/j.cca.2022.12.006
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。