文献速递临床生化临床免疫

HPLC-DMS/MS/MS法同时测定血清雄激素与17-羟孕酮及在PCOS中的应用

Sensitive HPLC-DMS/MS/MS method coupled with dispersive magnetic solid phase extraction followed by in situ derivatization for the simultaneous determination of multiplexing androgens and 17-hydroxyprogesterone in human serum and its application to patients with polycystic ovarian syndrome.
PubMed 文献 · Clin Chim Acta · 2022-11-28

该研究建立了一种基于分散磁固相萃取和原位衍生的HPLC-DMS/MS/MS方法,可同时测定人血清中7种雄激素和17-羟孕酮,并应用于432例临床样本。方法学验证显示精密度和准确度可接受,提取回收率54.0-92.7%,适用于多囊卵巢综合征患者的内源性雄激素检测。

要点

  • 该方法可同时测定7种雄激素和17-羟孕酮,适用于PCOS相关激素评估。
  • 方法学验证显示批内和批间精密度分别<16.7%和12.9%,准确度可接受。
  • 已应用于432例临床样本,分析性能优于传统电化学发光免疫法。

核心结论

  • 所建立的HPLC-DMS/MS/MS方法绿色、快速、灵敏且准确,可用于内源性雄激素测定,并易于在医学实验室推广。

研究设计与样本

  • 研究开发了HPLC-DMS/MS/MS方法,采用Fe3O4@GO纳米颗粒进行分散磁固相萃取,以Girard试剂P进行原位衍生化,并应用于432例临床样本的测定。

主要发现

  • 8种分析物的批内和批间精密度分别<16.7%和12.9%,相对误差分别为-14.7-13.3%和-9.3-11.0%。
  • 不同分析物的提取回收率为54.0-92.7%。
  • 该方法已应用于432例临床样本的测定。

关键数值与推荐

  • 线性范围根据临床用途设定。
  • 批内精密度<16.7%,批间精密度<12.9%。
  • 批内相对误差-14.7-13.3%,批间相对误差-9.3-11.0%。
  • 提取回收率54.0-92.7%。

临床怎么用

  • 该方法可为医学实验室提供优于传统电化学发光免疫法的分析性能,适用于PCOS患者内源性雄激素的准确量化。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND AND AIMS: Androgens play important roles in polycystic ovarian syndrome (PCOS). However, measures of androgens based on mass spectrometry (MS) remain complex due to endogenous inferences of isomers or compounds with similar structures. Lack of sensitivity can also affect the accurate quantification of androgens, especially for very low level of 11-oxygenated androgens. MATERIALS AND METHODS: We developed a fast and sensitive high-performance liquid chromatography-differential mobility spectrometry tandem mass spectrometry (HPLC-DMS/MS/MS) method for the simultaneous determination of seven androgens and 17-hydroxyprogesterone. Dispersive magnetic solid phase extraction (DMSPE) was conducted with core-shell structured nanoparticles of magnetic graphene oxide (Fe3O4@GO). In situ derivatization was performed using Girard's Reagent P. RESULTS: Linear ranges of the eight analytes were set in terms of clinical use. Intra- and inter-run precisions were < 16.7 % and 12.9 % for all the analytes and relative error was - 14.7-13.3 % and - 9.3-11.0 %, respectively. Extraction recoveries were 54.0-92.7 % for different analytes. The method was validated and was applied to assay 432 clinical samples. CONCLUSION: The developed method is green, fast, sensitive and accurate for the determination of endogenous androgens. It can be readily implemented in medical laboratories to provide superior analytical performance over the traditional electrochemiluminescence immunoassay method.

来源:Clin Chim Acta,查看原文。版权归原作者所有。

引用

  1. Zhang X, Xu H, Feng G, et al. Sensitive HPLC-DMS/MS/MS method coupled with dispersive magnetic solid phase extraction followed by in situ derivatization for the simultaneous determination of multiplexing androgens and 17-hydroxyprogesterone in human serum and its application to patients with polycystic ovarian syndrome. Clin Chim Acta. 2022. PMID: 36450312. doi:10.1016/j.cca.2022.11.025

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。