常规生化与血液学参数的移动均值质控:实施工具箱
该研究在实验室信息系统(LIS)中实施移动均值质控(MA QC),评估9个常规生化与血液学参数,其中3个因高变异和LIS技术问题被排除,6个显示对室内质控(IQC)有附加价值并实施,部分参数还用于趋势监测等。
要点
- MA QC对6个常规参数有附加价值,可补充传统IQC。
- Hb、ESR、钠因高变异或LIS问题未实施MA QC。
- MA QC尤其适用于趋势监测、维护后小偏移检测和仪器间比对。
核心结论
- 实施MA QC对实验室有附加价值,尤其适用于趋势监测、维护后小偏移检测和仪器间比对。
研究设计与样本
- 基于LIS数据提取,对9个常规生化与血液学参数进行MA QC有用性评估和设置优化,采用偏倚检测曲线。
主要发现
- 9个参数中3个(Hb、ESR、钠)因高变异和LIS技术问题被排除。
- 剩余6个参数MA QC显示对IQC有附加价值并已实施。
- 3个参数建立了直接的MA报警处理流程,包括LIS内置新功能。
- 对其他参数识别出超出实时监测的MA用途。
关键数值与推荐
- 评估参数包括血清钠、钙、碳酸氢盐、游离甲状腺素、血红蛋白、平均红细胞体积、平均红细胞血红蛋白浓度、网织红细胞计数和红细胞沉降率。
临床怎么用
- MA QC可用于趋势监测、检测维护后小偏移和仪器间比对。
- 实施前需评估参数变异和LIS技术可行性,优化MA QC设置。
局限性
- 部分参数因高变异和LIS技术问题未能实施MA QC。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Moving average quality control (MA QC) is a patient-based real-time quality control system. Advantages compared to conventional periodic internal quality control (IQC) include absence of commutability problems and continuous monitoring of performance. We implemented MA QC for multiple routine hematology and chemistry parameters. We describe the evaluation process and provide practical tools to aid MA QC implementation. METHODS: Nine parameters (serum sodium, calcium, bicarbonate and free thyroxine, hemoglobin [Hb], mean corpuscular volume, mean corpuscular hemoglobin concentration [MCHC], reticulocyte count and erythrocyte sedimentation rate [ESR]) were chosen for initial consideration. Using data extractions from the laboratory information system (LIS; General Laboratory Information Management System), evaluation of usefulness and optimization of MA QC settings was performed using bias detection curves. After this, MA QC settings were incorporated in our LIS for further evaluation and implementation in routine care. RESULTS: Three out of nine parameters (Hb, ESR, and sodium) were excluded from MA QC implementation due to high variation and technical issues in the LIS. For the six remaining parameters, MA QC showed added value to IQC and was therefore implemented in the LIS. For three parameters a direct MA alarm work-up method was set up, including newly developed built-in features in the LIS. For the other parameters, we identified MA utilization beyond real-time monitoring. CONCLUSIONS: Implementation of MA QC has added value for our laboratory setting. Additional utilization beyond real-time QC monitoring was identified. We find MA QC especially useful for trend monitoring, detection of small shifts after maintenance and inter-analyzer comparisons.
引用
- van Andel E, Henricks LM, Giliams APM, et al. Moving average quality control of routine chemistry and hematology parameters - a toolbox for implementation. Clin Chem Lab Med. 2022. PMID: 36044749. doi:10.1515/cclm-2022-0655
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。