超高效液相色谱-串联质谱法测定低血清睾酮的临床实用性

Clinical Usefulness of Ultraperformance Liquid Chromatography-Tandem Mass Spectrometry Method for Low Serum Testosterone Measurement.
PubMed 文献 · Ann Lab Med · 2022-09-01

该研究开发并验证了超高效液相色谱-串联质谱(UPLC-MS/MS)法测定血清总睾酮,评估精密度、准确度、线性、定量下限(LLOQ)、携带污染、离子抑制、稳定性和参考区间。结果显示该方法性能可接受,LLOQ为0.03 nmol/L,低于免疫分析法;在睾酮浓度<1.67 nmol/L时,Architect免疫分析法显示正百分比偏倚。该方法可用于低睾酮浓度的准确定量。

要点

  • UPLC-MS/MS法测定血清总睾酮的LLOQ为0.03 nmol/L,低于免疫分析法。
  • 在睾酮浓度<1.67 nmol/L时,Architect免疫分析法显示正百分比偏倚。
  • 该方法精密度和准确度可接受,血清睾酮在4°C、-20°C及三次冻融循环后稳定。

核心结论

  • UPLC-MS/MS法性能可接受,LLOQ低于免疫分析法,可用于低睾酮浓度的准确定量。

研究设计与样本

  • 开发并验证UPLC-MS/MS法测定血清总睾酮,评估精密度、准确度、线性、LLOQ、携带污染、离子抑制、稳定性和参考区间。
  • 方法比较:在40例睾酮浓度0.14-55.48 nmol/L的受试者及额外160份睾酮浓度<1.67 nmol/L的血清中,与Architect i2000免疫分析法比较。

主要发现

  • 批内和批间精密度CV<2.81%,准确度偏倚<3.85%,均可接受。
  • 验证线性区间为0.03-180.84 nmol/L,LLOQ为0.03 nmol/L。
  • 未观察到显著携带污染和离子抑制;血清睾酮在4°C、-20°C及三次冻融循环后稳定。
  • 参考区间成功验证;在0.14-55.48 nmol/L范围内相关性良好,但Architect法在<1.67 nmol/L时显示正百分比偏倚。

关键数值与推荐

  • 批内和批间精密度CV<2.81%。
  • 准确度偏倚<3.85%。
  • 验证线性区间0.03-180.84 nmol/L。
  • LLOQ为0.03 nmol/L。

临床怎么用

  • 对于低睾酮浓度的样本,UPLC-MS/MS法可提供比免疫分析法更低的定量下限和更准确的定量结果。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Mass spectrometry methods exhibit higher accuracy and lower variability than immunoassays at low testosterone concentrations. We developed and validated an ultraperformance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) assay for quantifying serum total testosterone. METHODS: We used an ExionLC UPLC (Sciex, Framingham, MA, USA) system and a Sciex Triple Quad 6500+ (Sciex) MS/MS system in electrospray ionization and positive ion modes with multiple reaction monitoring transitions to evaluate precision, accuracy, linearity, lower limit of quantitation (LLOQ), carryover, ion suppression, stability, and reference intervals. For method comparison, we measured serum testosterone concentrations using this method in 40 subjects whose testosterone concentrations ranged from 0.14 to 55.48 nmol/L as determined using the Architect i2000 immunoassay (Abbott Diagnostics, Abbott Park, IL, USA) and in an additional 160 sera with testosterone concentrations <1.67 nmol/L. RESULTS: The intra- and inter-run precision CVs were <2.81%, and the accuracy bias values were <3.85%, which were all acceptable. The verified linear interval was 0.03-180.84 nmol/L; the LLOQ was 0.03 nmol/L. No significant carryover and ion suppression were observed. The testosterone in serum was stable at 4°C, at -20°C, and after three freeze-thaw cycles. The reference intervals were successfully verified. The correlation was good at testosterone concentrations of 0.14-55.48 nmol/L; however, the Architect assay showed positive percent bias at concentrations <1.67 nmol/L. CONCLUSIONS: The UPLC-MS/MS assay shows acceptable performance, with a lower LLOQ than the immunoassay. This method will enable the quantitation of low testosterone concentrations.