文献速递临床微生物质量管理与危急值

MetaSystems自动化荧光显微镜系统辅助检测临床样本中抗酸杆菌的评估

Evaluation of MetaSystems Automated Fluorescent Microscopy System for the Machine-Assisted Detection of Acid-Fast Bacilli in Clinical Samples.
PubMed 文献 · J Clin Microbiol · 2022-09-19

本研究评估MetaSystems自动化荧光抗酸杆菌玻片扫描分析仪的准确性,以培养为参考,检测133份培养阳性与363份培养阴性呼吸道、组织、体液等样本。独立使用时敏感度97.0%、特异度12.7%;辅助技师阅片时敏感度70.7%、特异度89.0%。手工镜检敏感度79.7%、特异度98.6%,系统尚需改进方能用于临床辅助。

要点

  • 系统独立判读敏感度高(97.0%)但特异度极低(12.7%),假阳性多。
  • 辅助技师阅片时敏感度70.7%、特异度89.0%,均低于手工镜检。
  • 系统敏感度不受涂片分级或菌种影响,但11.5%玻片扫描失败。

核心结论

  • MetaSystems自动化荧光抗酸杆菌扫描分析仪目前性能尚不足以辅助临床实验室镜检人员,需进一步改进。

研究设计与样本

  • 以培养法为参考,评估133份培养阳性和363份培养阴性的临床样本(呼吸道284份、组织120份、体液81份、其他11份)的吖啶橙染色玻片,使用MetaSystems Mycobacteria Scanner运行NEON Metafer AFB模块。

主要发现

  • 系统未能扫描57份(11.5%)玻片。
  • 系统独立使用时敏感度97.0%(129/133;95% CI 92.5至99.2),特异度12.7%(46/363;95% CI 9.4至16.5)。
  • 系统辅助技师阅片时敏感度70.7%(94/133;95% CI 62.2至78.3),特异度89.0%(323/363;95% CI 85.3至92.0)。
  • 手工镜检敏感度79.7%(106/133;95% CI 71.9至86.2),特异度98.6%(358/363;95% CI 96.8至99.6)。
  • 系统敏感度不受涂片分级或分枝杆菌菌种影响;70.3%的假阳性涂片在系统判读中≥2+。

临床怎么用

  • 当前不建议将该系统用于临床实验室辅助镜检,需等待性能改进。

局限性

  • 系统未能扫描57份(11.5%)玻片,可能影响评估完整性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Manual reading of fluorescent acid-fast bacilli (AFB) microscopy slides is time-intensive and technically demanding. The aim of this study was to evaluate the accuracy of MetaSystems' automated fluorescent AFB slide scanner and analyzer. Auramine O-stained slides corresponding to 133 culture-positive and 363 culture-negative respiratory (n = 284), tissue (n = 120), body fluid (n = 81), and other (n = 11) sources were evaluated with the MetaSystems Mycobacteria Scanner running the NEON Metafer AFB Module. The sensitivity and specificity of the MetaSystems platform was measured as a standalone diagnostic and as an assistant to technologists to review positive images. Culture results were used as the reference method. The MetaSystems platform failed to scan 57 (11.5%) slides. The MetaSystems platform used as a standalone had a sensitivity of 97.0% (129/133; 95% CI 92.5 to 99.2) and specificity of 12.7% (46/363; 95% CI 9.4 to 16.5). When positive scans were used to assist technologists, the MetaSystems platform had a sensitivity of 70.7% (94/133; 95% CI 62.2 to 78.3) and specificity of 89.0% (323/363; 95% CI 85.3 to 92.0). The manual microscopy method had a sensitivity of 79.7% (106/133; 95% CI 71.9 to 86.2) and specificity of 98.6% (358/363; 95% CI 96.8 to 99.6). The sensitivity of the MetaSystems platform was not impacted by smear grade or mycobacterial species. The majority (70.3%) of false positive smears had ≥2+ smear results with the MetaSystems platform. Further performance improvements are needed before the MetaSystems' automated fluorescent AFB slide reader can be used to assist microscopist in the clinical laboratory.

来源:J Clin Microbiol,查看原文。版权归原作者所有。

引用

  1. Tomasello G, Foroughi F, Padron D, et al. Evaluation of MetaSystems Automated Fluorescent Microscopy System for the Machine-Assisted Detection of Acid-Fast Bacilli in Clinical Samples. J Clin Microbiol. 2022. PMID: 36121216. doi:10.1128/jcm.01131-22

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。