同型半胱氨酸参考物质互通性评估
研究采用五种常规方法同时测定30份血清和7种同型半胱氨酸参考物质,以液相色谱串联质谱为参考方法,按CLSI EP30-A和IFCC文件评估互通性。EP30-A下7种参考物质均通过评估,35种组合中34种可互通;IFCC下28种组合结论明确,7种无法判定。结论认为同型半胱氨酸参考物质互通性良好。
要点
- EP30-A评估下7种同型半胱氨酸参考物质均通过互通性评估。
- 五种常规方法与7种参考物质的35种组合中34种可互通。
- IFCC评估下28种组合结论明确,7种组合无法判定。
核心结论
- 同型半胱氨酸参考物质显示出良好的互通性。
- 测量程序的灵敏度、测量偏差与不确定度及不同方法所选「被测量」的差异可能影响评估结果。
- 不同方法采用不同判定标准可解释评估结果的差异。
研究设计与样本
- 采用五种常规测量方法同时测定30份血清样本和7种同型半胱氨酸参考物质(来自美国国家标准与技术研究院和中国计量科学研究院)。
- 以液相色谱串联质谱法作为参考方法。
- 依据CLSI EP30-A和IFCC互通性评估文件两种方法评估7种参考物质的互通性。
主要发现
- 按EP30-A评估,7种参考物质均通过互通性评估,五种常规方法与7种参考物质的35种组合中34种可互通。
- 按IFCC评估,28种组合的互通性评估结论明确(可互通或不可互通),其余7种组合结果无法判定。
关键数值与推荐
- EP30-A评估:7种参考物质通过,35种组合中34种可互通。
- IFCC评估:28种组合结论明确,7种组合无法判定。
临床怎么用
- 同型半胱氨酸参考物质互通性良好,可用于常规方法校准与结果溯源。
- 评估参考物质互通性时需注意测量程序灵敏度、偏差与不确定度及方法间「被测量」差异的影响。
局限性
- 不同方法采用不同判定标准可能导致评估结果出现差异。
- IFCC评估中有7种组合的互通性结果无法判定。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Commutability of reference materials is essential for ensuring the traceability of patient measurement results and the technical basis for the use of reference materials. Commutability is only relevant for matrixed reference material; it is a prerequisite for the accuracy and authenticity of calibration methods. In this study, we evaluated the commutability of reference materials for homocysteine. METHODS: Five conventional measurement methods were applied to simultaneously measure 30 serum samples and seven homocysteine reference materials from the National Institute of Standards and Technology and the National Institute of Metrology. Liquid chromatography tandem-mass spectrometry was used as a reference method. Two methods were used to evaluate the commutability of the seven reference materials according to the Clinical and Laboratory Standards Institute EP30-A and the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) commutability assessment document. RESULTS: Among 35 combinations of the five conventional methods and seven reference materials, after evaluation in accordance with the EP30-A, the seven reference materials passed the commutability assessment, and 34 combinations were commutable. According to the IFCC, the commutability evaluation of 28 combinations was conclusive (commutable or non-commutable), while results for the remaining seven combinations could not be determined. CONCLUSIONS: The homocysteine reference materials showed good commutability. The sensitivity of the measurement procedure, measurement deviation and uncertainty, and differences in the "measurand" selected by different methods may affect the evaluation results. Additionally, different judgment standards for different methods may explain the observed variations in evaluation results.