研究提出基于患者数据的机器学习实时质量控制(MLiQC),对中国计量院标准参考数据可溯源,并与IFCC推荐的四种PBRTQC方法比较。

要点

  • MLiQC基于随机森林建立,可溯源至中国计量院标准参考数据。
  • 在ALB关键偏倚下,MLiQC不确定度为0.14%。
  • MLiQC的ANPed平均从600降至20,优于四种最优PBRTQC。

核心结论

  • MLiQC在临床有效性上高度优于PBRTQC,适合真实临床场景。
  • 从算法溯源和临床有效性两方面验证,模型性能令人满意。

研究设计与样本

  • 数据处理后划分独立训练集和测试集,用随机森林建立MLiQC。
  • 通过计量学算法溯源和IFCC分析工作组推荐的四种PBRTQC方法进行验证。

主要发现

  • 以ALB关键偏倚为例,MLiQC不确定度为0.14%,由中国计量院标准参考数据评估。
  • 与四种最优PBRTQC方法在关键偏倚下比较,MLiQC的ANPed平均从600降至20。
  • 所有检测项目的MLiQC的MNPed和95NPed均优于上述最优PBRTQC 90%以上。

关键数值与推荐

  • ALB关键偏倚下MLiQC不确定度为0.14%。
  • MLiQC的ANPed平均从600降至20。
  • MLiQC的MNPed和95NPed在所有项目上优于最优PBRTQC 90%以上。

临床怎么用

  • MLiQC可作为内部质量控制的替代或整合工具用于真实临床场景。
  • 模型可溯源至中国计量院标准参考数据,支持临床有效性评估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Patient-based real-time quality control (PBRTQC) has gained attention as an alternative/integrative tool for internal quality control (iQC). However, it is still doubted for its performance and its application in real clinical settings. We aim to generate a newly and easy-to-access patient-based real-time QC by machine learning (ML) traceable to standard reference data with assigned values by National Institute of Metrology of China (NIM), and to compare it with PBRTQC for clinical validity evaluation. METHODS: For five representative biochemistry analytes, 1,195 000 patient testing results each were collected. After data processing, independent training and test sets were divided. Machine learning internal quality control (MLiQC) was set up by Random Forest in ML and was validated by way of both metrology algorithm traceability and 4 PBRTQC methods recommended by IFCC analytical working group. RESULTS: MLiQC were established. As an example of albumin (ALB) at the critical bias, the uncertainty of MLiQC was 0.14%, which was evaluated by standard reference data produced by NIM. Compared with four optimal PBRTQC methods at critical bias, the average of the number of patient samples from a bias introduced until detected (ANPed) of MLiQC averagely decreased from 600 to 20. The median and 95 quantiles of NPeds (MNPed and 95NPed) of MLiQC were superior to all optimal PBRTQCs above 90% for all test items. CONCLUSIONS: MLiQC is highly superior to PBRTQC and well-suited in real settings. The validation of the model from two aspects of algorithm traceability and clinical effectiveness confirms its satisfactory performance.

来源:Clin Chem Lab Med,查看原文。版权归原作者所有。

引用

  1. Zhou R, Wang W, Padoan A, et al. Traceable machine learning real-time quality control based on patient data. Clin Chem Lab Med. 2022. PMID: 35852126. doi:10.1515/cclm-2022-0548

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。