文献速递临床生化质量管理与危急值

多项式回归与加权最小二乘回归在分析测量范围验证中的比较

Comparison between polynomial regression and weighted least squares regression analysis for verification of analytical measurement range.
PubMed 文献 · Clin Chem Lab Med · 2022-05-09

该研究在三个实验室用五个样本、每样本两个重复,对临床化学生化项目的分析测量范围(AMR)进行验证,并分别用多项式回归与加权最小二乘法(WLS)解读同一批数据。多项式回归下42.3-56.8%的化学项目通过线性验证,而WLS下为63.5-78.3%。作者认为按新修订的CLSI EP6-ED2采用WLS解读AMR验证数据可减少实验室工作量。

要点

  • 同一AMR验证数据用WLS解读时通过线性验证的项目比例高于多项式回归。
  • WLS下所有五个样本偏离线性的90%置信区间落在允许偏离线性(ADL)内或与ADL重叠的比例较高。
  • 按CLSI EP6-ED2采用WLS解读可减少实验室工作量。

核心结论

  • 按新修订的CLSI EP6-ED2用WLS方法解读AMR验证数据,可减少实验室工作量并实现高效实验室实践。

研究设计与样本

  • 在三个不同实验室对临床化学检验项目进行AMR验证,使用五个样本、每样本两个重复。
  • 对每个实验室的同一评价数据分别用多项式回归分析和WLS方法分析后比较结果。

主要发现

  • 多项式回归分析下三个实验室分别有42.3-56.8%的化学项目线性得到验证。
  • 同一数据用WLS分析时63.5-78.3%的检验项目线性得到验证,且所有五个样本的偏离线性均在ADL标准内。
  • 所有偏离线性的90%置信区间与ADL重叠的情况为78.8-91.3%。

关键数值与推荐

  • 线性验证判定:多项式回归分析与WLS方法分别判断五个样本浓度范围内线性是否得到验证。
  • WLS判定条件之一为偏离线性的90%置信区间是否包含在允许偏离线性(ADL)内。

临床怎么用

  • 临床实验室可按CLSI EP6-ED2采用WLS解读AMR验证数据,以减少验证工作量。
  • 解读时需关注偏离线性的90%置信区间与ADL的关系。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Recently, the linearity evaluation protocol by the Clinical & Laboratory Standards Institute (CLSI) has been revised from EP6-A to EP6-ED2, with the statistical method of interpreting linearity evaluation data being changed from polynomial regression to weighted least squares linear regression (WLS). We analyzed and compared the analytical measurement range (AMR) verification results according to the present and prior linearity evaluation guidelines. METHODS: The verification of AMR of clinical chemistry tests was performed using five samples with two replicates in three different laboratories. After analyzing the same evaluation data in each laboratory by the polynomial regression analysis and WLS methods, results were compared to determine whether linearity was verified across the five sample concentrations. In addition, whether the 90% confidence interval of deviation from linearity by WLS was included in the allowable deviation from linearity (ADL) was compared. RESULTS: A linearity of 42.3-56.8% of the chemistry items was verified by polynomial regression analysis in three laboratories. For analysis of the same data by WLS, a linearity of 63.5-78.3% of the test items was verified where the deviation from linearity of all five samples was within the ADL criteria, and the cases where the 90% confidence interval of all deviation from linearity overlapped the ADL was 78.8-91.3%. CONCLUSIONS: Interpreting AMR verification data by the WLS method according to the newly revised CLSI document EP6-ED2 could reduce laboratory workload, enabling efficient laboratory practice.

来源:Clin Chem Lab Med,查看原文。版权归原作者所有。

引用

  1. Jeong TD, Kim SK, Kim S, et al. Comparison between polynomial regression and weighted least squares regression analysis for verification of analytical measurement range. Clin Chem Lab Med. 2022. PMID: 35531706. doi:10.1515/cclm-2022-0018

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。