HarmoSter研究:校准对LC-MS/MS检测循环皮质醇、17OH-孕酮和醛固酮可比性的影响

Report from the HarmoSter study: impact of calibration on comparability of LC-MS/MS measurement of circulating cortisol, 17OH-progesterone and aldosterone.
PubMed 文献 · Clin Chem Lab Med · 2022-02-16

该研究比较了9个欧洲实验室采用LC-MS/MS检测皮质醇、17OH-孕酮和醛固酮的性能,并评估了外部质量评估(EQA)材料和校准的影响。使用本地校准(CAL1)时,实验室间中位CV分别为4.9%、11.8%和13.8%;采用外部校准(CAL3)后降至3.6%、10.3%和8.6%(均p<0.001)。采用通用可溯源校准品和RMP定值的EQA材料有助于LC-MS/MS类固醇检测的标准化。

要点

  • 本地校准下,皮质醇、17OH-孕酮和醛固酮的实验室间中位CV分别为4.9%、11.8%和13.8%。
  • 采用外部校准CAL3后,三者实验室间中位CV分别降至3.6%、10.3%和8.6%(均p<0.001)。
  • CAL3改善了皮质醇和17OH-孕酮的实验室间偏倚和相关性,但残余变异仍存在。

核心结论

  • 采用通用可溯源校准品和RMP定值的EQA材料有益于LC-MS/MS类固醇检测的标准化,但残余变异仍存在。

主要发现

  • 使用CAL1时,皮质醇、17OH-孕酮和醛固酮的实内CV范围分别为2.8-7.4%、4.4-18.0%和5.2-22.2%。
  • 使用CAL1时,皮质醇、17OH-孕酮和醛固酮的实验室间中位CV分别为4.9%、11.8%和13.8%;使用CAL3时分别为3.6%、10.3%和8.6%(均p<0.001)。
  • 使用CAL1时,与所有实验室中位数的中位偏倚范围分别为皮质醇-6.6至6.9%、17OH-孕酮-17.2至7.8%、醛固酮-12.0至16.8%。
  • 使用CAL3改善了皮质醇和17OH-孕酮的实验室间方法偏倚和相关性。
  • EQA材料的正确度和偏倚大多可接受,但部分材料存在不合适的互换性和靶值赋值问题。

关键数值与推荐

  • 使用CAL1时,皮质醇、17OH-孕酮和醛固酮的实内CV范围分别为2.8-7.4%、4.4-18.0%和5.2-22.2%。
  • 使用CAL1时,皮质醇、17OH-孕酮和醛固酮的实验室间中位CV分别为4.9%、11.8%和13.8%;使用CAL3时分别为3.6%、10.3%和8.6%(均p<0.001)。

临床怎么用

  • LC-MS/MS检测类固醇时,采用通用可溯源校准品可改善实验室间可比性。
  • EQA材料应关注互换性和靶值赋值,以确保评估有效性。

局限性

  • 部分EQA材料存在不合适的互换性和靶值赋值问题。
  • 使用CAL1时,大多数实验室的回归线显著偏离最佳拟合线。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Liquid chromatography-tandem mass spectrometry (LC-MS/MS) is recommended for measuring circulating steroids. However, assays display technical heterogeneity. So far, reproducibility of corticosteroid LC-MS/MS measurements has received scant attention. The aim of the study was to compare LC-MS/MS measurements of cortisol, 17OH-progesterone and aldosterone from nine European centers and assess performance according to external quality assessment (EQA) materials and calibration. METHODS: Seventy-eight patient samples, EQA materials and two commercial calibration sets were measured twice by laboratory-specific procedures. Results were obtained by in-house (CAL1) and external calibrations (CAL2 and CAL3). We evaluated intra and inter-laboratory imprecision, correlation and agreement in patient samples, and trueness, bias and commutability in EQA materials. RESULTS: Using CAL1, intra-laboratory CVs ranged between 2.8-7.4%, 4.4-18.0% and 5.2-22.2%, for cortisol, 17OH-progesterone and aldosterone, respectively. Trueness and bias in EQA materials were mostly acceptable, however, inappropriate commutability and target value assignment were highlighted in some cases. CAL2 showed suboptimal accuracy. Median inter-laboratory CVs for cortisol, 17OH-progesterone and aldosterone were 4.9, 11.8 and 13.8% with CAL1 and 3.6, 10.3 and 8.6% with CAL3 (all p<0.001), respectively. Using CAL1, median bias vs. all laboratory-medians ranged from -6.6 to 6.9%, -17.2 to 7.8% and -12.0 to 16.8% for cortisol, 17OH-progesterone and aldosterone, respectively. Regression lines significantly deviated from the best fit for most laboratories. Using CAL3 improved cortisol and 17OH-progesterone between-method bias and correlation. CONCLUSIONS: Intra-laboratory imprecision and performance with EQA materials were variable. Inter-laboratory performance was mostly within specifications. Although residual variability persists, adopting common traceable calibrators and RMP-determined EQA materials is beneficial for standardization of LC-MS/MS steroid measurements.