基层医疗即时检验内部质量控制频率评分系统的开发
本研究基于系统文献综述和Noklus全国性POC组织的共识流程,开发了一套评分系统,用于确定基层医疗中不同POCT设备应多久进行一次内部质量控制(IQC)。该系统考虑分析物重要性、设备类型、用户友好性和患者样本量四个因素,并应用于17种分析物和134种设备。结果显示,大多数检测高风险分析物的设备(74种中的71种)获得每日或每周IQC频率推荐,例如所有血细胞计数仪应每日进行IQC,所有血糖仪应每周进行IQC。
要点
- 评分系统基于分析物重要性、设备类型和用户友好性三项因素之和确定IQC频率。
- 患者样本量决定是否及如何调整个体诊所的IQC频率。
- 检测高风险分析物的设备中,71/74被推荐每日或每周IQC。
核心结论
- 本研究提出了一种基于共识的评分系统,可为基层医疗POCT设备提供差异化、设备特定的IQC频率推荐。
研究设计与样本
- 基于系统文献综述和涉及Noklus全国性POC组织的完整流程,开发评分系统,并应用于17种分析物和134种POCT设备(153种分析物-设备组合)。
主要发现
- 大多数检测高风险分析物的设备(74种中的71种)获得每日或每周IQC频率推荐。
- 所有血细胞计数仪应每日进行IQC,所有血糖仪应每周进行IQC。
关键数值与推荐
- 评分系统考虑四个因素:分析物在诊断和监测中的重要性、POCT设备类型、用户友好性、患者样本量。
- 前三项因素评分之和决定一般推荐的IQC频率,患者样本量决定是否及如何调整个体诊所的频率。
临床怎么用
- 该评分系统易于采用至其他本地环境,使用简便,可帮助基层医疗确定POCT设备的IQC频率。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Internal quality control (IQC) plays an important role in quality assurance in laboratory medicine. However, there is no universal consensus or guideline on when and how IQC should be analyzed on point-of-care testing (POCT) devices. The aim of this study was to develop a scoring system to determine how often IQC should be analyzed in primary healthcare on the various POCT devices. METHODS: Based on a systematic literature review and a thorough process involving the whole Noklus, a nationwide POC organization, a scoring system for when to analyze IQC was developed. Four factors were considered to significantly impact IQC frequency: The importance of the analyte in diagnosing and monitoring patients, type of POCT device, user-friendliness, and number of patient samples. For each POCT device, the first three factors were given a score, and the sum of the scores determined the general recommended IQC frequency. The number of patient samples determined whether and how to adjust these frequencies in each individual general practice. RESULTS: The scoring system was applied to 17 analytes and 134 different POCT devices (153 analyte-device combinations). Most of the devices analyzing high-risk analytes (71 out of 74) obtained daily or weekly IQC frequency. For example, all blood-cell counters and all glucose meters should undergo IQC daily and weekly, respectively. CONCLUSIONS: This study presents a consensus-based scoring system for differentiated and device-specific recommendations for IQC frequency on POCT devices in primary healthcare. The scoring system can easily be adopted to other local environments and is easy to use.
引用
- Gidske G, Sandberg S, Fossum AL, et al. Point-of-care testing in primary healthcare: a scoring system to determine the frequency of performing internal quality control. Clin Chem Lab Med. 2022. PMID: 35150123. doi:10.1515/cclm-2021-1258
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。