文献速递临床免疫临床生化

IgG4相关疾病患者中的IgG4-IgE复合物

IgG4-IgE complex in a patient with IgG4-related disease.
PubMed 文献 · Clin Chim Acta · 2022-01-20

本文报告一例81岁男性IgG4相关疾病(IgG4-RD)患者,其IgE浓度在不同检测试剂间存在显著差异(Elecsys IgE II为1287.0 IU/mL,Iatroace IgE为60.9 IU/mL),IgG4浓度为675 mg/dL。通过蛋白G和抗IgG4抗体纯化及尺寸排阻HPLC分析,发现IgG4与IgE结合形成复合物,导致试剂间IgE检测结果不一致。

要点

  • IgG4-RD患者中IgE升高需警惕IgG4-IgE复合物导致的检测差异。
  • 不同IgE检测试剂结果不一致时,应考虑IgG4-IgE复合物干扰。
  • IgG4-RD中IgE产生增加可能具有临床意义。

核心结论

  • IgG4-RD患者中IgG4与IgE结合形成复合物,可导致不同IgE试剂检测结果不一致。

研究设计与样本

  • 病例报告:一例81岁男性IgG4-RD伴组织学自身免疫性胰腺炎患者,采用蛋白G和抗IgG4抗体偶联基质纯化及尺寸排阻HPLC分析。

主要发现

  • Elecsys IgE II检测IgE为1287.0 IU/mL,Iatroace IgE检测为60.9 IU/mL,IgG4浓度为675 mg/dL。
  • IgG纯化流出液和结合 fractions 的IgE浓度分别为6.8 IU/mL(10.8%)和56.2 IU/mL(89.2%)。
  • IgG4纯化流出液和结合 fractions 的IgE浓度分别为6.8 IU/mL(12.2%)和49.0 IU/mL(87.8%)。
  • 尺寸排阻HPLC中IgE洗脱为单一峰(640 kDa),IgG4洗脱显示小峰(640 kDa)和大峰(170 kDa)。

临床怎么用

  • 当IgG4-RD患者IgE检测结果与临床不符或不同试剂差异大时,应考虑IgG4-IgE复合物可能。

局限性

  • 为单例病例报告,结论需更大样本验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: IgE concentrations are occasionally elevated in patients with IgG4-related disease (IgG4-RD). In this report, we describe a novel case of IgG4-RD in which IgE concentrations were discordant between measuring reagents. CASE: An 81-year-old man was diagnosed with IgG4-RD and histological autoimmune pancreatitis, which ensued without treatment. The IgE concentrations measured using Elecsys IgE II Immunoassay and Iatroace IgE were 1287.0 IU/mL and 60.9 IU/mL, respectively. IgG4 concentration was 675 mg/dL. METHODS: To identify IgG and IgG4 directly bound to IgE, purification using protein G and anti-IgG4 antibody-conjugated matrixes and size-exclusion high-performance liquid chromatography (HPLC) were performed. RESULTS: In purification analysis, the IgE concentration of the flow-through and bound fractions were 6.8 IU/mL (10.8%) and 56.2 IU/mL (89.2%) for IgG purification and 6.8 IU/mL (12.2%) and 49.0 IU/mL (87.8%) for IgG4 purification. IgE was eluted as a single peak (640 kDa) using size-exclusion HPLC. In the elution pattern of IgG4, a minor peak (640 kDa) and a major peak (170 kDa) were observed. These results indicate that IgG4 binds to IgE and forms a complex, resulting in a discrepancy between reagents. CONCLUSIONS: In this report, we present an IgG4-IgE complex in a patient with IgG4-RD, which affected the discrepancy in IgE concentrations between IgE reagents. This report points to the significance of increased IgE production in IgG4-RD.

来源:Clin Chim Acta,查看原文。版权归原作者所有。

引用

  1. Nakano K, Sugita J, Mafune N, et al. IgG4-IgE complex in a patient with IgG4-related disease. Clin Chim Acta. 2022. PMID: 35065908. doi:10.1016/j.cca.2022.01.013

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。