文献速递临床生化临床免疫

无抗体靶向质谱法测定血清嗜铬粒蛋白A

Antibody-Free Quantification of Serum Chromogranin A by Targeted Mass Spectrometry.
PubMed 文献 · Clin Chem · 2021-01-01

与CISBIO免疫测定法比较200份血清标本,Pearson相关系数r=0.953,CgA值平均高2-4倍,两法一致性80.9%(95%CI 72.8%-89.2%)。截短和翻译后修饰未直接导致两法结果不一致。

要点

  • 与免疫测定法相关性好(r=0.953),但CgA值平均高2-4倍,一致性80.9%。
  • 截短和翻译后修饰(含2个新磷酸化位点)未直接导致两法结果不一致。

核心结论

  • LC-MS/MS定量CgA提供了分析灵敏且可重复的替代商业免疫测定法的方法。

研究设计与样本

  • 方法开发与验证研究,使用混合模式阴离子交换固相萃取板从血清提取CgA,胰蛋白酶消化后LC-MS/MS分析,并与CISBIO免疫测定法比较200份既往送检血清标本。

主要发现

  • 与CISBIO免疫测定法Pearson相关系数r=0.953,CgA值平均高2-4倍。
  • 两法一致性80.9%(95%CI 72.8%-89.2%),显示高度一致。
  • 截短和翻译后修饰(包括2个此前未观察或预测的磷酸化位点)未直接导致两法结果不一致。

关键数值与推荐

  • 回收率89%-115%。
  • 批内和批间不精密度<10%。
  • 与CISBIO免疫测定法Pearson相关系数r=0.953。
  • 两法一致性80.9%(95%CI 72.8%-89.2%)。

临床怎么用

  • LC-MS/MS法可作为CgA免疫测定法的替代方法,具有更宽的动态范围,无需稀释高浓度标本。

局限性

  • 与免疫测定法相比CgA值平均高2-4倍,两法结果不能直接互换。
  • 截短和翻译后修饰未直接导致不一致,但其他因素可能影响一致性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Chromogranin A (CgA) is a 48 kDa protein that serves as a diagnostically sensitive, but nonspecific, serum biomarker for neuroendocrine tumors. Immunoassays for CgA are not standardized and have a narrow dynamic range, which requires dilution of concentrated specimens. We developed and validated an antibody-free, liquid chromatography-tandem mass spectrometry (LC-MS/MS)-based method for CgA without these limitations. METHODS: CgA was extracted from serum using a mixed-mode anion exchange solid-phase extraction plate, digested with trypsin, and analyzed by LC-MS/MS using well-characterized CgA calibration standards. After validation, the mass spectrometry method was compared with the CISBIO immunoassay using 200 serum specimens previously submitted for CgA analysis. Specimens with discordant results were reanalyzed by high-resolution mass spectrometry- (HRMS) -based methods to assess the contribution of truncated and post-translationally modified forms of CgA. RESULTS: The assay had a linear range of 50 to 50 000 ng/mL, recoveries between 89% and 115%, and intra- and interassay imprecision <10%. LC-MS/MS assay results showed a Pearson's correlation of r = 0.953 with the CISBIO immunoassay, with CgA values being a mean 2- to 4-fold higher. Concordance for CgA between the 2 assays was 80.9% (95% CI 72.8%-89.2%), showing substantial agreement. Truncation and posttranslational modification, including 2 phosphorylation sites that had not been previously observed or predicted to our knowledge, did not appear to contribute directly to discordance between the 2 assays. CONCLUSION: Quantification of CgA by LC-MS/MS provides an analytically sensitive and reproducible alternative to commercially available immunoassays.

来源:Clin Chem,查看原文。版权归原作者所有。

引用

  1. Weber DM, Yang JY, Goldman SM, et al. Antibody-Free Quantification of Serum Chromogranin A by Targeted Mass Spectrometry. Clin Chem. 2021. PMID: 34718463. doi:10.1093/clinchem/hvab191

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。