四种多重实时PCR试剂盒检测泌尿生殖道及性传播病原体的临床表现

Clinical performance of four multiplex real-time PCR kits detecting urogenital and sexually transmitted pathogens.
PubMed 文献 · Clin Microbiol Infect · 2021-10-02

该研究以425份样本(主要为尿液及宫颈阴道、咽、直肠拭子)评估四种多重实时PCR试剂盒检测沙眼衣原体、淋病奈瑟菌、生殖支原体和阴道毛滴虫的性能,并与cobas CT/NG和TV/MG试剂盒比较。结果显示四种试剂盒对沙眼衣原体检测性能良好,但对生殖支原体和阴道毛滴虫的阳性符合率较低(分别为63.3-74.1%和51.2-68.4%),Seegene和Sacace试剂盒对淋病奈瑟菌的阳性符合率也较低(分别为71.2%和63.1%)。

要点

  • 四种多重实时PCR试剂盒检测沙眼衣原体性能良好。
  • 生殖支原体和阴道毛滴虫阳性符合率低,范围为63.3-74.1%和51.2-68.4%。
  • 不应将解脲脲原体和微小脲原体等定植菌纳入此类试剂盒。

核心结论

  • 多重实时PCR试剂盒可一步检测多种性传播感染相关病原体,但部分试剂盒对生殖支原体和阴道毛滴虫的检测性能较弱。

研究设计与样本

  • 研究使用425份样本(主要为尿液及宫颈阴道、咽、直肠拭子),将四种多重实时PCR试剂盒与cobas CT/NG和TV/MG试剂盒及in-house TaqMan PCR进行比较。

主要发现

  • 四种试剂盒检测沙眼衣原体性能良好。
  • 与cobas MG/TV试剂盒相比,四种试剂盒检测生殖支原体和阴道毛滴虫的阳性符合率范围分别为63.3-74.1%和51.2-68.4%。
  • Seegene和Sacace试剂盒检测淋病奈瑟菌的阳性符合率分别为71.2%(95%CI 61.8-79.0)和63.1%(95%CI 53.5-71.8)。
  • ELITechGroup试剂盒检测淋病奈瑟菌的阴性符合率为92.5%(95%CI 89.1-94.9),Fast-Track试剂盒检测生殖支原体的阴性符合率为93.2%(95%CI 89.6-95.7),略低于其他试剂盒。

临床怎么用

  • 用户应意识到部分试剂盒对生殖支原体和阴道毛滴虫的检测性能较弱。
  • 此类试剂盒不应包含解脲脲原体和微小脲原体等人定植菌的检测。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: We evaluated the clinical performances of four multiplex real-time PCR commercial kits for the detection of Chlamydia trachomatis, Neisseria gonorrhoeae, Mycoplasma genitalium and Trichomonas vaginalis: the STI PLUS ELITe MGB kit (ELITechGroup), N. gonorrhoeae/C. trachomatis/M. genitalium/T.vaginalis Real-TM kit (Sacace Biotechnologies), Allplex STI Essential kit (Seegene), and FTD Urethritis Plus kit (Fast-Track Diagnostics). METHODS: The kit performance for C. trachomatis, N. gonorrhoeae, M. genitalium and T. vaginalis detection was compared to that of the cobas CT/NG and TV/MG kits (Roche Diagnostics) using 425 samples, mainly urine and cervicovaginal, throat and rectal swabs. Detection of Ureaplasma parvum, U. urealyticum and Mycoplasma hominis were compared to that of in-house TaqMan PCRs. RESULTS: The four kits showed good performances for the detection of C. trachomatis. They all presented a low positive agreement for the detection of M. genitalium and T. vaginalis (ranges 63.3-74.1% and 51.2-68.4%, respectively) compared to the cobas MG/TV kit. The Seegene and Sacace kits showed additional low positive agreement for the detection of N. gonorrhoeae (71.2%, 95%CI 61.8-79.0 and 63.1%, 95%CI 53.5-71.8, respectively). We observed a slight but significant lower negative agreement for N. gonorrhoeae detection using the ELITechGroup kit (92.5%, 89.1-94.9) and for M. genitalium detection using the Fast-Track kit (93.2%, 89.6-95.7) compared to other kits. CONCLUSION: Multiplex real-time PCR kits are convenient methods for the detection of several pathogens associated with sexually transmitted infections (STIs) in a single step, but colonizing Ureaplasma spp. and M. hominis species should not be included in these kits. Users should be aware of the weak performance of some kits for the detection of M. genitalium and T. vaginalis.

来源:Clin Microbiol Infect,查看原文。版权归原作者所有。

引用

  1. Pereyre S, Caméléna F, Hénin N, et al. Clinical performance of four multiplex real-time PCR kits detecting urogenital and sexually transmitted pathogens. Clin Microbiol Infect. 2021. PMID: 34610459. doi:10.1016/j.cmi.2021.09.028

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。