文献速递新发突发、热带与寄生虫病

人工智能辅助自动显微镜用于疟疾诊断:多中心研究

Artificial Intelligence-Supported Automated Microscopy for Malaria Diagnosis: A Multicenter Study.
PubMed 文献 · Clin Infect Dis · 2026-09-30

该多中心研究在埃塞俄比亚和加纳采集2201例发热患者样本,评估Noul miLab自动显微镜(整合涂片制备、染色、成像和AI辅助寄生虫检测)对疟疾的诊断性能。以专家显微镜和qPCR为金标准,miLab对恶性疟原虫和间日疟原虫的灵敏度均高,且特异性良好,灵敏度显著优于卫生中心本地显微镜检查。

要点

  • miLab对恶性疟原虫灵敏度96.3%(专家镜检金标准)和97.4%(qPCR>200/µL金标准)。
  • miLab对间日疟原虫灵敏度96.8%(专家镜检金标准)和95.9%(qPCR>200/µL金标准)。
  • miLab特异性对比qPCR:恶性疟98.8%,间日疟97.8%。

核心结论

  • miLab自动显微镜对恶性疟原虫和间日疟原虫诊断具有高灵敏度和特异性。
  • miLab灵敏度显著优于卫生中心本地显微镜检查。

研究设计与样本

  • 多中心研究,在埃塞俄比亚2个点和加纳1个点采集2201例发热患者样本。
  • 样本同时用本地显微镜、miLab和快速诊断试验筛查,以qPCR和专家显微镜为金标准。

主要发现

  • 以专家显微镜为金标准,miLab诊断恶性疟原虫灵敏度96.3%(335/348),间日疟原虫灵敏度96.8%(399/412)。
  • 以qPCR阳性且密度>200寄生虫/µL为金标准,miLab诊断恶性疟原虫灵敏度97.4%(298/306),间日疟原虫灵敏度95.9%(419/437)。
  • 对比qPCR,miLab特异性恶性疟原虫98.8%,间日疟原虫97.8%。
  • 在埃塞俄比亚,miLab阳性样本中正确鉴定虫种的比例:恶性疟原虫单感染98.7%,间日疟原虫单感染96.2%。

临床怎么用

  • miLab可作为发热患者疟疾诊断的自动化工具,减少镜检者间差异。
  • 在恶性疟和间日疟流行区,miLab的高灵敏度支持其用于临床诊断和监测。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Accurate malaria diagnosis is key for patient management, surveillance, and control. Automated microscopy can overcome the variation observed among microscopists and is a promising new tool for diagnosis. The Noul miLab integrates smear preparation, staining, imaging, and artificial intelligence-supported parasite detection in a portable device. METHODS: A total of 2201 samples were collected from febrile patients across 2 sites in Ethiopia, where Plasmodium falciparum and Plasmodium vivax are frequent, and in Ghana, where P. falciparum transmission is intense. Samples were screened using local microscopy at the health center, miLab, and rapid diagnostic tests. Quantitative polymerase chain reaction (qPCR) and expert microscopy were used as gold standards. RESULTS: The miLab reached a sensitivity for P. falciparum diagnosis of 96.3% (335 of 348) using expert microscopy as the gold standard and a sensitivity of 97.4% (298 of 306) using qPCR-positive infections at densities >200 parasites/µL as the gold standard. The sensitivity of miLab for P. vivax was 96.8% (399 of 412) using expert microscopy as the gold standard and 95.9% (419 of 437) using qPCR-positive infections at densities >200 parasites/µL as the gold standard. Compared to qPCR, specificity was 98.8% for P. falciparum and 97.8% for P. vivax. miLab was significantly more sensitive than microscopy conducted at the health center. In Ethiopia, among miLab-positive samples, miLab assigned the correct parasite species to 98.7% of P. falciparum and 96.2% of P. vivax mono-infections. CONCLUSIONS: The miLab automated microscope shows high sensitivity and specificity for P. falciparum and P. vivax diagnosis.

来源:Clin Infect Dis,查看原文。版权归原作者所有。

引用

  1. Hawaria D, Ewnetu Y, Kamugisha H, et al. Artificial Intelligence-Supported Automated Microscopy for Malaria Diagnosis: A Multicenter Study. Clin Infect Dis. 2026. PMID: 42814025. doi:10.1093/cid/ciag484