快速多粘菌素试验检测革兰阴性菌多粘菌素耐药性的诊断准确性:系统综述与荟萃分析
汇总灵敏度为0.966(95%CI 0.92-0.98),特异度为0.945(95%CI 0.92-0.96),AUC为0.979。肠杆菌目结果稳定且极重大误差率低,非发酵菌数据有限且异质性大,建议实施时结合确认试验。
要点
- 快速多粘菌素试验汇总灵敏度0.966、特异度0.945,AUC 0.979,诊断准确性高。
- 肠杆菌目灵敏度与特异度均高且极重大误差率低,非发酵菌数据有限需谨慎。
- 灵敏度存在显著异质性(I2=86.1%),非发酵菌高流行区应结合确认试验。
核心结论
- 快速多粘菌素试验对革兰阴性菌多粘菌素耐药检测具有高诊断准确性,尤其适用于肠杆菌目。
- 非发酵菌数据有限且异质性显著,实施时应结合确认试验。
研究设计与样本
- 纳入评估快速多粘菌素试验对比肉汤微量稀释法的研究,对象为革兰阴性菌分离株。
- 采用双变量随机效应模型汇总灵敏度和特异度,并按菌种类型进行亚组分析。
- 使用QUADAS-2工具评估偏倚风险,PROSPERO注册号CRD420251140151。
主要发现
- 汇总灵敏度为0.966(95%CI 0.92-0.98),汇总特异度为0.945(95%CI 0.92-0.96)。
- AUC为0.979,提示高诊断准确性。
- 亚组分析显示肠杆菌目与非发酵菌之间存在菌种特异性差异。
- 灵敏度存在显著无法解释的异质性(I2=86.1%)。
- 肠杆菌目中灵敏度和特异度持续较高,极重大误差率低。
临床怎么用
- 快速多粘菌素试验可用于肠杆菌目多粘菌素耐药的快速检测。
- 在非发酵革兰阴性杆菌高流行环境中,实施快速试验应伴随确认试验。
- 非发酵菌结果解释需谨慎,因数据有限且异质性大。
局限性
- 灵敏度存在显著无法解释的异质性(I2=86.1%)。
- 非发酵菌数据有限,影响该亚组的结论可靠性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Polymyxins are considered last-resort agents for treating multidrug-resistant Gram-negative infections; however, resistance is increasing worldwide. Broth microdilution is the standard reference method for colistin susceptibility testing; however, its technical complexity and prolonged turnaround time have prompted the development of the rapid polymyxin test. The clinical performance of these rapid tests across different bacterial species remains unclear. OBJECTIVES: To evaluate the diagnostic accuracy of the rapid polymyxin test for detecting polymyxin resistance in Gram-negative bacteria. METHODS: Data sources: PubMed, Scopus, Web of Science, and Embase (inception to 31 August 2025). STUDY ELIGIBILITY CRITERIA: Studies evaluating rapid polymyxin tests against broth microdilution. PARTICIPANTS: Gram-negative bacterial isolates. INDEX TEST: Rapid polymyxin-based assays. REFERENCE STANDARD: Broth microdilution. ASSESSMENT OF RISK OF BIAS: Quality Assessment of Diagnostic Accuracy Studies 2 tool. METHODS OF DATA SYNTHESIS: Bivariate random-effects model for summary sensitivity and specificity; hierarchical summary receiver operating characteristic curves; subgroup analyses by organism type. PROSPERO: CRD420251140151. RESULTS: Thirty-seven studies were included. The summary sensitivity was 0.966 (95% CI, 0.92-0.98), and the specificity was 0.945 (95% CI, 0.92-0.96). The area under the curve was 0.979, indicating high diagnostic accuracy. Subgroup analyses revealed species-specific variations, particularly between Enterobacterales and nonfermenters. Substantial unexplained heterogeneity was present (I2 = 86.1%) for sensitivity. Among Enterobacterales, sensitivity and specificity were consistently high, with low very major error rates. DISCUSSION: The rapid polymyxin test demonstrates high diagnostic accuracy, particularly for Enterobacterales, but substantial heterogeneity and limited data for nonfermenters warrant caution. Implementation should be accompanied by confirmatory testing, especially in settings with a high prevalence of nonfermenting Gram-negative bacilli.
引用
- C V J, Reddy Varikasuvu S, C P H, et al. Diagnostic accuracy of the rapid polymyxin test for colistin resistance in Gram-negative bacteria: a systematic review and meta-analysis. Clin Microbiol Infect. 2026. PMID: 42601019. doi:10.1016/j.cmi.2026.08.009
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。