肝素诱导的血小板减少症实验室诊断的国际外部质量评估

An international external quality assessment for laboratory diagnosis of heparin-induced thrombocytopenia.
PubMed 文献 · J Thromb Haemost · 2019-02-20

由德国格赖夫斯瓦尔德大学医学院输血医学科组织首次国际外部质量评估,向德国、加拿大和新加坡的七家参与者分发六份患者血清和一份阴性对照,评估PF4/肝素IgG酶免疫分析与功能试验。PF4/肝素IgG试验定性符合率为97.6%,功能试验定性符合率明显较低,除阴性对照外所有样本达到共识。

要点

  • PF4/肝素IgG酶免疫分析的定性符合率为97.6%。
  • 功能试验定性符合率明显较低,阴性对照出现非特异性反应。
  • 所有HIT阴性样本均被正确诊断,HIT阳性样本共识度达70%。

核心结论

  • 联合酶免疫分析和功能试验的外部质量评估是可行的,但样本材料有限是障碍。
  • 功能试验的定性符合率明显低于PF4/肝素IgG试验,需要开展功能试验的外部质量评估。

研究设计与样本

  • 由德国格赖夫斯瓦尔德大学医学院输血医学科组织的试点国际外部质量评估。
  • 向德国、加拿大和新加坡的七家参与者分发六份转诊至格赖夫斯瓦尔德进行HIT诊断的患者血清和一份阴性对照样本。
  • 参与者报告本地检测PF4/肝素复合物IgG特异性抗体的酶免疫分析光密度值以及功能试验(HIPA或SRA)结果。
  • 共识定义为至少70%一致,即七家参与实验室中至少五家一致。

主要发现

  • 七家参与者中六家报告了酶免疫分析结果,定量符合率为97.6%。
  • 功能试验除阴性对照外所有样本达到共识,阴性对照中部分参与者报告非特异性反应。
  • 所有HIT阴性样本均被所有参与者正确诊断。
  • HIT阳性样本达到70%共识。

局限性

  • 样本材料有限是需克服的障碍。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Essentials A pilot study for External Quality Assessment for testing of HIT is described. The qualitative accordance for the PF4/heparin IgG test was 97.6%. The qualitative accordance for functional HIT tests was considerably lower. External Quality Assessment for functional HIT tests is required. SUMMARY: Objective Heparin-induced thrombocytopenia (HIT) is a potentially life-threatening complication of heparin exposure. Diagnosis is most reliable using a combination of an enzyme immunoassay (EIA) that detects antibodies against platelet factor 4 (PF4)/heparin complexes ("antigen" assay) and a "functional" assay that detects platelet-activating properties of the pathogenic HIT antibodies. No External Quality Assessment (EQA) is available for a combination of the tests. Here we report on the results of the first international EQA. Methods The pilot EQA was organized by the Department of Transfusion Medicine, Universitätsmedizin Greifswald, Germany. Six serum samples of patients, which were referred to Greifswald for HIT diagnosis, and one negative control sample were distributed to seven participants in Germany, Canada, and Singapore. Participants were asked to report the optical density (OD) values of their local EIA test for IgG-specific antibodies against the PF4/heparin complexes and the results for a functional assay (HIPA or SRA). Consensus was defined as a minimum 70% agreement, i.e., agreement among at least five of the seven participating laboratories. Results and conclusion Six out of seven participants reported results for EIA, with a high quantitative accordance (97.6%). For the functional assay, consensus was reached for all samples except the negative control, for which some participants reported nonspecific reactivity. All HIT-negative samples were correctly diagnosed by all participants; for HIT-positive samples, consensus of 70% was reached. Although the limited availability of sample material is an obstacle to overcome, an EQA combining both EIA and functional testing is feasible.

来源:J Thromb Haemost,查看原文。版权归原作者所有。