Myoton量化硬化性慢性移植物抗宿主病的重现性与重复性
该研究纳入36例硬化性慢性移植物抗宿主病成人患者,由2名研究人员在2次研究访视中,使用Myoton设备在7个双侧部位(14个解剖部位)测量软组织振荡频率和松弛时间。结果显示观察者内和观察者间的整体测量一致性和可靠性良好,但单个部位的可重复性不如患者整体平均值。
要点
- Myoton测量硬化性cGVHD皮肤特性的观察者间ICC>0.90,观察者内ICC为0.85和0.84。
- 整体观察者间平均成对绝对差<5%,观察者内标准化平均成对绝对差<7%。
- 患者整体平均测量比单个部位测量更可重复,支持在患者层面使用Myoton。
核心结论
- Myoton用于评估硬化性cGVHD皮肤特性具有良好重现性和重复性,尤其适用于患者层面的整体测量。
研究设计与样本
- 纳入36例硬化性慢性移植物抗宿主病成人患者,由2名研究人员在2次研究访视中,使用Myoton设备在7个双侧部位(14个解剖部位)测量振荡频率和松弛时间。
主要发现
- 两个Myoton参数的整体观察者间平均成对绝对差均小于平均整体值的5%,观察者间ICC均大于0.90。
- 两次访视的中位间隔为47.5天。
- 两个Myoton参数的整体标准化观察者内平均成对绝对差均小于平均整体值的7%。
- 频率和松弛时间的观察者内ICC分别为0.85和0.84。
- 每次研究访视选定的附加部位的重现性和重复性与标准14个解剖部位相似。
临床怎么用
- 作者强调Myoton用于评估硬化性cGVHD皮肤特性时,应采用患者层面的整体测量。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
There is an urgent need for validated tools to measure sclerotic cutaneous chronic graft-versus-host disease (scGVHD). We examined the interobserver reproducibility within a session and intraobserver repeatability between sessions of the Myoton device for quantifying skin sclerosis in 36 adults with scGVHD. The Myoton was used to measure oscillation frequency and relaxation time of soft tissues at 7 bilateral sites (14 anatomic sites) by 2 study personnel at 2 study sessions. Agreement was measured using mean pairwise absolute difference (MPD), and reliability was measured using intraclass correlation coefficient (ICC). For each of the 2 Myoton parameters, the overall interobserver MPD was <5% of the average overall values and the interobserver ICC was >0.90 between the 2 observers, indicating excellent agreement and reliability within a measurement session. The median time between sessions 1 and 2 was 47.5 days. The overall normalized intraobserver MPD was <7% of the average overall values for each of the 2 Myoton parameters, reflecting good agreement between sessions. The intraobserver ICC for frequency and relaxation time parameters were 0.85 and 0.84, respectively, indicating good reliability between sessions. The reproducibility and repeatability of a bonus site selected at each study visit were similar to the standard 14 anatomic sites. However, no individual site was nearly as reproducible or repeatable as the overall Myoton measurements averaged across the patient. Our findings emphasize the utility of the Myoton for assessing skin properties in scGVHD with patient-level measurements.
引用
- Farhadfar N, El Jurdi N, Baker KK, et al. Reproducibility and repeatability of the Myoton to quantify sclerotic chronic graft-versus-host disease. Blood Adv. 2026. PMID: 41191519. doi:10.1182/bloodadvances.2025016648