双次高水平消毒与液化学灭菌对ERCP十二指肠镜再处理的比较:一项前瞻性随机研究
该前瞻性随机研究于2017年10月23日至2018年9月24日比较双次高水平消毒(DHLD)与液化学灭菌(LCS)对ERCP十二指肠镜再处理的效果。共878份再处理后监测培养(DHLD组453份,LCS组425份),总体阳性17份(1.9%);DHLD组阳性8份(1.8%)与LCS组阳性9份(2.1%)无显著差异(P=.8)。两组高关注度微生物培养均2份(.5% vs .5%,P=1.0),未检出多重耐药菌。结论为两种方法阳性培养率均低,但均未完全消除阳性培养。
要点
- DHLD组阳性培养8份(1.8%),LCS组9份(2.1%),无显著差异(P=.8)。
- 两组高关注度微生物培养均2份(.5% vs .5%,P=1.0)。
- 未检出包括耐碳青霉烯肠杆菌科在内的多重耐药菌。
核心结论
- DHLD与LCS两种再处理策略的阳性培养率均低,但均未完全消除十二指肠镜的阳性培养。
研究设计与样本
- 前瞻性随机研究,于2017年10月23日至2018年9月24日比较DHLD与LCS两种十二指肠镜再处理方式,按内部方案对再处理后十二指肠镜随机采样培养监测。
- 共纳入878份再处理后监测培养,其中DHLD组453份,LCS组425份。
主要发现
- 全部培养中17份为任何微生物阳性(1.9%)。
- DHLD组阳性培养8份(1.8%),LCS组阳性培养9份(2.1%),两组无显著差异(P=.8)。
- 两组高关注度微生物培养均为2份(DHLD组.5% vs LCS组.5%,P=1.0)。
- 未检出多重耐药菌,包括耐碳青霉烯肠杆菌科。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND AND AIMS: The potential for transmission of pathogenic organisms is a problem inherent to the current reusable duodenoscope design. Recent outbreaks of multidrug-resistant pathogenic organisms transmitted via duodenoscopes has brought to light the urgency of this problem. Microbiologic culturing of duodenoscopes and reprocessing with repeat high-level disinfection (HLD) or liquid chemical sterilization (LCS) have been offered as supplemental measures to enhance duodenoscope reprocessing by the U.S. Food and Drug Administration. This study aims to compare the efficacy of reprocessing duodenoscopes with double HLD (DHLD) versus LCS. METHODS: We prospectively evaluated 2 different modalities of duodenoscope reprocessing from October 23, 2017 to September 24, 2018. Eligible duodenoscopes were randomly segregated to be reprocessed by either DHLD or LCS. Duodenoscopes were randomly cultured after reprocessing for surveillance based on an internal protocol. RESULTS: During the study period, there were 878 post-reprocessing surveillance cultures (453 in the DHLD group and 425 in the LCS group). Of all cultures, 17 were positive for any organism (1.9%). There was no significant difference of positive cultures when comparing the duodenoscopes undergoing DHLD (8 positive cultures, 1.8%) with duodenoscopes undergoing LCS (9 positive cultures, 2.1%; P = .8). Both groups had 2 cultures that grew high-concern organisms (.5% vs .5%, P=1.0). No multidrug-resistant organisms, including carbapenem-resistant enterobacteriaceae, were detected. CONCLUSIONS: DHLD and LCS both resulted in a low rate of positive cultures, for all organisms and for high-concern organisms. However, neither process completely eliminated positive cultures from duodenoscopes reprocessed with 2 different supplemental reprocessing strategies.
引用
- Gromski MA, Sieber MS, Sherman S, et al. Double high-level disinfection versus liquid chemical sterilization for reprocessing of duodenoscopes used for ERCP: a prospective randomized study. Gastrointest Endosc. 2020. PMID: 32745532. doi:10.1016/j.gie.2020.07.057
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。