文献速递食管胃疾病与幽门螺杆菌消化内镜

用于区分食管疾病的黏膜阻抗轮廓分析系统的开发与验证

Development and Validation of a Mucosal Impedance Contour Analysis System to Distinguish Esophageal Disorders.
PubMed 文献 · Gastroenterology · 2019-01-31

该前瞻性研究纳入69例接受食管胃十二指肠镜的患者,开发球囊黏膜阻抗(MI)导管系统,按GERD(n=24)、EoE(n=21)和非GERD(n=24)分组,发现三组食管轴向MI模式差异显著(P<.01),并在3家独立机构的数据中验证了MI模式与诊断的关联。

要点

  • 球囊MI导管可在内镜下即时检测食管黏膜完整性变化。
  • 非GERD患者各测量节段MI值均较高;GERD远端MI低而近端正常化,EoE各节段MI均低。
  • 开发队列中MI截距和斜率识别GERD的AUC=0.67、EoE的AUC=0.84、非GERD的AUC=0.83。

核心结论

  • 研究开发了球囊MI导管系统,可在内镜下即时检测食管黏膜完整性变化,发现其安全并能识别GERD、EoE或非GERD患者。
  • 研究在独立队列中验证了上述发现。

研究设计与样本

  • 前瞻性研究纳入69例接受食管胃十二指肠镜(伴或不伴无线pH监测)的患者,分为GERD组(糜烂性食管炎或pH异常;n=24)、EoE组(经远端和近端食管组织病理分析确认;n=21)和非GERD组(食管胃十二指肠镜和pH检测结果正常;n=24)。
  • 使用受试者工作特征曲线和曲线下面积(AUC)比较球囊MI的诊断准确性,采用多项逻辑回归估计各组归属概率,并用3家独立机构患者的数据验证MI模式与诊断的关联。

主要发现

  • GERD、EoE和非GERD患者的食管轴向MI模式差异显著(P<.01)。
  • 非GERD患者所有测量节段的MI值均较高。
  • GERD患者远端食管MI值低、近端食管正常化;EoE患者食管所有节段MI值均低。
  • 开发队列中,距鳞柱交界距离增加时的MI截距和上升速率(斜率)识别GERD的AUC=0.67、EoE的AUC=0.84、非GERD的AUC=0.83。
  • 1例患者发生不良事件(术后轻度胸痛),出院后无进一步事件。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND & AIMS: Diagnostic testing for chronic esophageal disorders relies on histopathology analysis of biopsies or uncomfortable transnasal catheters or wireless pH monitoring, which capture abnormal intraluminal refluxate. We therefore developed a balloon mucosal impedance (MI) catheter system that instantly detects changes in esophageal mucosal integrity during endoscopy over a long segment of the esophagus. We performed a prospective study to evaluate the ability of a balloon-incorporated MI catheter to detect and evaluate esophageal disorders, including gastroesophageal reflux disease (GERD) and eosinophilic esophagitis (EoE). METHODS: We performed a prospective study of 69 patients undergoing esophagogastroduodenoscopy with or without wireless pH monitoring. Patients were classified as having GERD (erosive esophagitis or abnormal pH; n = 24), EoE (confirmed with pathology analysis of tissues from both distal and proximal esophagus; n = 21), or non-GERD (normal results from esophagogastroduodenoscopy and pH tests; n = 24). Receiver operating characteristic curves and area under the operating characteristic curve (AUC) were used to compare the accuracy of balloon MI in diagnosis. Probabilities of assignment to each group (GERD, non-GERD, or EoE) were estimated using multinomial logistic regression. Association between MI patterns and diagnoses were validated using data from patients seen at 3 separate institutions. RESULTS: MI pattern along the esophageal axis differed significantly (P < .01) among patients with GERD, EoE, and non-GERD. Patients with non-GERD had higher MI values along all measured segments. The MI pattern for GERD was easily distinguished from that of EoE: in patients with GERD, MI values were low in the distal esophagus and normalized along the proximal esophagus, whereas in patients with EoE, measurements were low in all segments of the esophagus. Intercept and rate of rise of MI value (slope) as distance increased from the squamocolumnar junction identified patients with GERD with an AUC = 0.67, patients with EoE with an AUC = 0.84, and patients with non-GERD with an AUC = 0.83 in the development cohort. One patient had an adverse event (reported mild chest pain after the procedure) and was discharged from the hospital without further events. CONCLUSIONS: We developed a balloon MI catheter system that instantly detects changes in esophageal mucosal integrity during endoscopy and found it to be safe and able to identify patients with GERD, EoE, or non-GERD. We validated our findings in a separate cohort for patients. ClinicalTrials.gov ID NCT03103789.

来源:Gastroenterology,查看原文。版权归原作者所有。