无线动力胶囊在疑似胃轻瘫患者中诊断与性能特征的验证
前瞻性研究纳入10个中心167例疑似胃轻瘫受试者(53例糖尿病,114例非糖尿病),2013至2016年同步行无线动力胶囊(WMC)与胃排空闪烁显像(GES)检查。WMC检出胃排空延迟的比例(34.6%)高于GES(24.5%)(P=.009),两者总体一致率75.7%(kappa=0.42)。WMC还检出GES未发现的胃肠区域及弥漫性转运异常(61.8%受试者)。
要点
- WMC检出胃排空延迟比例34.6%,高于GES的24.5%(P=.009)。
- 非糖尿病受试者中WMC检出延迟比例33.3%,高于GES的17.1%(P<.001)。
- GES检出快速胃排空比例13.8%,高于WMC的3.3%(P<.001)。
核心结论
- WMC与GES对胃排空的分析存在一致性,但WMC诊断产出高于GES,能更频繁检出胃排空延迟并识别胃外转运异常。
- 糖尿病与非糖尿病受试者的GES与WMC结果不同,这些发现可能影响疑似胃轻瘫患者的管理。
研究设计与样本
- 前瞻性研究,纳入10个中心的167例疑似胃轻瘫受试者(53例糖尿病,114例非糖尿病),2013年至2016年同步行GES与WMC检查。
主要发现
- WMC检出胃排空延迟的比例为34.6%,高于GES的24.5%(P=.009)。
- 两种方法总体一致率为75.7%(kappa=0.42)。
- 非糖尿病受试者中,WMC检出胃排空延迟的比例为33.3%,高于GES的17.1%(P<.001)。
- 糖尿病受试者中GES检出胃排空延迟的比例为41.7%,高于非糖尿病受试者的17.1%(P=.002)。
- WMC检出严重胃排空延迟的比例为13.8%,高于GES的6.9%(P=.02);GES检出快速胃排空的比例为13.8%,高于WMC的3.3%(P<.001)。
关键数值与推荐
- GES定义胃排空延迟为4小时食物滞留超过10%,WMC定义为胶囊进入十二指肠时间超过5小时。
- 严重胃排空延迟定义为WMC胃排空时间超过12小时或GES 4小时滞留超过35%。
- 快速胃排空定义为GES 1小时食物滞留低于38%或WMC胃排空时间低于1小时45分钟。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND & AIMS: It is a challenge to make a diagnosis of gastroparesis. There is good agreement in results from wireless motility capsule (WMC) analysis and gastric emptying scintigraphy (GES), but the diagnostic yield of WMC is unclear and the accuracy of this method has not been validated. We compared the performance characteristics of WMC vs GES in assessing gastric emptying in patients with suspected gastroparesis. METHODS: We performed a prospective study of 167 subjects with gastroparesis (53 with diabetes and 114 without) at 10 centers, from 2013 through 2016. Subjects were assessed simultaneously by GES and with a WMC to measure gastric emptying and regional transit. Delayed gastric emptying by GES was defined as more than 10% meal retention at 4 hrs whereas delayed gastric emptying by WMC was defined as more than 5 hrs for passage of the capsule into the duodenum; a severe delay in gastric emptying was defined as a gastric emptying time of more than 12 hrs by WMC or more than 35% retention at 4 hrs by GES. Rapid gastric emptying was defined as less than 38% meal retention at 1 hr based on by GES or gastric emptying times less than 1:45 hrs by WMC. We compared diagnostic and performance characteristics of GES vs WMC. RESULTS: Delayed gastric emptying was detected in a higher proportion of subjects by WMC (34.6%) than by GES (24.5%) (P=.009). Overall agreement in results between methods was 75.7% (kappa=0.42). In subjects without diabetes, the WMC detected a higher proportion of subjects with delayed gastric emptying (33.3%) than GES (17.1%) (P < .001). A higher proportion of subjects with diabetes had delayed gastric emptying detected by GES (41.7%) compared with non-diabetic subjects (17.1%) (P=.002). Severe delays in gastric emptying were observed in a higher proportion of subjects by WMC (13.8%) than by GES (6.9%) (P = .02). Rapid gastric emptying was detected in a higher proportion of subjects by GES (13.8%) than by WMC (3.3%) (P < .001). Regional and generalized transit abnormalities were observed in 61.8% subjects and only detected by WMC. CONCLUSION: Although there is agreement in analysis of gastric emptying by GES vs WMC, WMC provides higher diagnostic yield than GES. WMC detects delayed gastric emptying more frequently than GES and identifies extra-gastric transit abnormalities. Diabetic vs non-diabetic subjects have different results from GES vs WMC. These findings could affect management of patients with suspected gastroparesis. ClinicalTrials.gov no: NCT02022826.
引用
- Lee AA, Rao S, Nguyen LA, et al. Validation of Diagnostic and Performance Characteristics of the Wireless Motility Capsule in Patients With Suspected Gastroparesis. Clin Gastroenterol Hepatol. 2018. PMID: 30557741. doi:10.1016/j.cgh.2018.11.063
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。