CC100结肠胶囊内镜系统的诊断性能与可行性:前瞻性多中心研究
在中国重庆3家三级医院开展的前瞻性多中心诊断准确性研究,共136例完成CC100结肠胶囊内镜和常规结肠镜。CC100完整结肠可视化率91.18%,与结肠镜总体检出一致率83.82%(κ=0.673);≥6 mm临床相关息肉灵敏度86.67%、特异度98.63%,无胶囊滞留或操作相关不良事件。
要点
- CC100完整结肠可视化率达91.18%,无胶囊滞留或操作相关不良事件。
- 与结肠镜总体检出一致率83.82%(κ=0.673)。
- ≥6 mm临床相关息肉灵敏度86.67%、特异度98.63%。
核心结论
- 新一代CC100结肠胶囊内镜系统可达到可接受的检查完整性和具有临床意义的诊断一致性。
研究设计与样本
- 前瞻性多中心诊断准确性研究,在中国重庆3家三级医院开展。
- 共136例参与者完成CC100结肠胶囊内镜后再行常规结肠镜作为参考标准。
主要发现
- CC100完整结肠可视化124/136例(91.18%),常规结肠镜完成136/136例(100%)。
- 与结肠镜总体检出一致率83.82%(κ=0.673)。
- ≥6 mm临床相关息肉灵敏度86.67%、特异度98.63%。
- 无胶囊滞留或操作相关不良事件发生。
临床怎么用
- 在常规结肠镜可及性受限场景下,CC100可作为微创筛查替代选择,但需结合其91.18%的完整可视化率评估检查失败风险。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND AND AIMS: Colorectal cancer screening continues to rely on conventional colonoscopy, yet its invasiveness, limited accessibility, and growing procedural demand constrain large-scale implementation. Colon capsule endoscopy (CCE) offers a minimally invasive alternative, but its clinical adoption has been hampered by incomplete examinations and concerns regarding diagnostic accuracy. METHODS: In this prospective multicenter diagnostic accuracy study conducted at 3 tertiary hospitals in Chongqing, China, participants underwent CCE (CC100, Jinshan Science and Technology Co Ltd, Chongqing, China) CCE followed by conventional colonoscopy as the reference standard. Primary outcomes included complete colonic visualization and diagnostic agreement for colorectal lesion detection. Secondary outcomes assessed lesion-specific performance and safety. RESULTS: Among 136 participants completing both procedures, complete colonic visualization was achieved with CC100 CCE in 124 of 136 participants (91.18%), whereas conventional colonoscopy was completed in 136 of 136 participants (100%). No capsule retention or procedure-related adverse events occurred. Overall agreement with colonoscopy detection was 83.82% (κ = 0.673). For clinically relevant polyps ≥6 mm, CC100 demonstrated a sensitivity of 86.67% and a specificity of 98.63%. CONCLUSIONS: This prospective multicenter study demonstrates that a next-generation CCE system can achieve acceptable examination completeness and clinically meaningful diagnostic concordance with colonoscopy.
引用
- Xie X, Guo J, Chen J, et al. Diagnostic performance and feasibility of the CC100 colon capsule endoscopy system: a prospective multicenter study. Gastrointest Endosc. 2026. PMID: 42497983. doi:10.1016/j.gie.2026.07.018
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。