肠道超声采用床旁与中心判读评估克罗恩病活动度的可靠性
本研究通过2天工作坊对24例克罗恩病患者进行肠道超声检查,6名当地判读者与4名中心判读者盲法评估8项参数及总体疾病活动度。5项参数(肠壁厚度、彩色多普勒信号、炎性肠系膜脂肪、黏膜下层突出、受累肠段长度)在当地与中心判读中均达到至少中等信度(ICC≥0.41),中心判读信度通常优于当地判读。
要点
- 肠壁厚度、彩色多普勒信号、炎性肠系膜脂肪、黏膜下层突出和受累肠段长度达到至少中等信度(ICC≥0.41)。
- 中心判读的可靠性通常优于当地判读。
- 肠壁分层和炎性肠系膜脂肪以二分类评估时ICC最高。
核心结论
- 当地与中心判读肠道超声对多项参数显示至少中等判读者间和判读者内信度。
- 本研究支持完善现有活动度指数并将肠道超声中心判读纳入临床试验。
研究设计与样本
- 前瞻性执行扫描,24例克罗恩病患者与6名胃肠病学家参与为期2天的工作坊,每例受试者共接受6次肠道超声扫描。
- 6名当地判读者与4名中心胃肠病学家-超声医师盲法评估8项肠道超声参数及总体超声疾病活动度。
- 信度采用组内相关系数(ICC)量化。
- 机构审查委员会批准本研究(12938)。
主要发现
- 5项肠道超声参数在当地与中心判读中均达到至少中等(ICC≥0.41)的判读者间和判读者内信度:肠壁厚度、彩色多普勒信号、炎性肠系膜脂肪、黏膜下层突出和受累肠段长度。
- 中心判读的可靠性通常优于当地判读。
- 肠壁分层和炎性肠系膜脂肪以二分类结局评估时ICC最高。
- 敏感性分析显示,肠道超声参数在评估最严重受累肠段时最可靠。
- 当地判读者识别最严重受累肠段时观察到中等信度。
临床怎么用
- 作者认为本研究支持完善现有活动度指数并将肠道超声中心判读纳入临床试验。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND & AIMS: Intestinal ultrasound (IUS) is increasingly used to assess Crohn's disease (CD) activity in clinical practice. However, application in clinical trials has been limited by heterogeneous scoring methods and concerns about reliability. We aimed to determine the inter- and intra-rater reliability of locally and centrally read IUS parameters for evaluating CD using prospectively performed scans. METHODS: Twenty-four participants with CD and 6 gastroenterologists participated in a 2-day workshop where each participant underwent 6 IUS scans in total. Eight IUS parameters (bowel wall thickness [BWT], bowel wall stratification [BWS], color Doppler signal [CDS], inflammatory mesenteric fat [i-fat], submucosal prominence, submucosal layer thickness, haustra coli/peristalsis, and affected segment length) and an overall measure of sonographic disease activity were blindly assessed by the 6 local readers and 4 central gastroenterologist-sonographers. Reliability was quantified using intraclass correlation coefficients (ICCs). Institutional review board approval was granted for this study (12938). RESULTS: Five IUS parameters demonstrated at least moderate (ICC ≥0.41) inter- and intra-rater reliability when local and central reading was performed (BWT, CDS, i-fat, submucosal prominence, and affected segment length). Reliability was generally better with central, in distinction to local, reading. ICCs for BWS and i-fat were highest when evaluated as binary outcomes. Sensitivity analyses demonstrated that IUS parameters are most reliable when evaluated in the worst affected segment. Fair reliability was observed when local readers identified the worst affected segment. CONCLUSIONS: Local and central reading of IUS demonstrated at least moderate inter- and intra-rater reliability for several parameters. This study supports refining existing activity indices and incorporating IUS central reading into clinical trials.
引用
- Goodsall TM, An YK, Andrews JM, et al. Reliability of Intestinal Ultrasound for Evaluating Crohn's Disease Activity Using Point-of-care and Central Reading. Clin Gastroenterol Hepatol. 2024. PMID: 39209200. doi:10.1016/j.cgh.2024.08.007