第二代结肠胶囊内镜柔性光谱成像色彩增强对结直肠息肉和肿瘤诊断效能的前瞻性研究
本研究前瞻性评估第二代结肠胶囊内镜(CCE)搭载柔性光谱成像色彩增强(FICE)对结直肠息肉和肿瘤(CRT)的诊断效能,以光学结肠镜为金标准,比较白光图像(CCE-WL)与FICE图像(CCE-FICE)的诊断准确性。纳入89例患者(75例有CRT),结果显示CCE-FICE在整体CRT、6-9mm及≥6mm CRTs的患者检出率及整体、<6mm、6-9mm、稍隆起、管状腺瘤、锯齿状息肉的病灶敏感度均显著高于CCE-WL,但≥10mm及隆起型CRT的敏感度差异无统计学意义;两种模式均检出全部腺癌。
要点
- CCE-FICE对整体CRT患者检出率88% vs CCE-WL 79%(P=0.0001),6-9mm者94% vs 63%(P=0.0055)。
- CCE-FICE病灶敏感度整体79% vs 61%(P<0.0001),<6mm者69% vs 53%(P<0.0001)。
- ≥10mm及隆起型CRT的敏感度差异无统计学意义(P=0.1138、P=0.0614)。
核心结论
- 与CCE-WL相比,CCE-FICE提高了除腺癌、≥10mm息肉和隆起型息肉外所有CRT的检出率。
研究设计与样本
- 前瞻性研究,纳入89例患者(65男24女),CCE后4个月内行光学结肠镜,两名结肠镜医师独立判读CCE-WL与CCE-FICE图像,以结肠镜所见为金标准比较诊断准确性。
主要发现
- 患者层面检出率:整体CRT CCE-WL 79% vs CCE-FICE 88%(P=0.0001);6-9mm CRT 63% vs 94%(P=0.0055);≥6mm CRT 78% vs 93%(P=0.0159)。
- 病灶层面敏感度:整体CRT CCE-WL 61% vs CCE-FICE 79%(P<0.0001);<6mm 53% vs 69%(P<0.0001);6-9mm 65% vs 93%(P=0.0007)。
- 病灶层面敏感度:稍隆起CRT 53% vs 75%(P<0.0001);管状腺瘤61% vs 79%(P<0.0001);锯齿状息肉57% vs 74%(P=0.0022)。
- ≥10mm CRT敏感度CCE-WL 81% vs CCE-FICE 94%(P=0.1138),隆起型CRT 77% vs 86%(P=0.0614),差异均无统计学意义。
- 两种模式均检出全部腺癌;整体CRT的Kappa系数CCE-WL为0.66、CCE-FICE为0.64,提示高度一致。
关键数值与推荐
- 75例有CRT者中包括36例锯齿状病变、63例腺瘤、9例腺癌。
临床怎么用
- 原文未据此给出临床推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND AND AIMS: We prospectively determined the efficacy of flexible spectral imaging color enhancement (FICE) used with second-generation colon capsule endoscopy (CCE) for colorectal polyps and tumors (CRTs). METHODS: This study included optical colonoscopy within 4 months after CCE. Two colonoscopists independently reviewed CCE using white-light images (CCE-WL) and CCE using FICE images (CCE-FICE), respectively. Based on colonoscopic findings as the criterion standard, the diagnostic accuracy for CRTs was compared between CCE-WL and CCE-FICE. RESULTS: Of 89 enrolled patients (65 men and 24 women; 75 with CRTs including 36 with serrated lesions, 63 with adenomas, and 9 with adenocarcinomas), the per-patient detectability of CCE-FICE for the representative CRTs was significantly higher than that of CCE-WL: overall CRTs (CCE-WL, 79%; CCE-FICE, 88%; P = .0001), 6- to 9-mm CRTs (CCE-WL, 63%; CCE-FICE, 94%; P = .0055), and ≥6-mm CRTs (CCE-WL, 78%; CCE-FICE, 93%; P = .0159). The per-lesion sensitivity of CCE-FICE was significantly higher than that of CCE-WL for CRTs: overall (CCE-WL, 61%; CCE-FICE, 79%; P < .0001), <6 mm (CCE-WL, 53%; CCE-FICE, 69%; P < .0001), 6- to 9-mm CRTs (CCE-WL, 65%; CCE-FICE, 93%; P = .0007), slightly elevated CRTs (CCE-WL, 53%; CCE-FICE, 75%; P < .0001), tubular adenomas (CCE-WL, 61%; CCE-FICE, 79%; P < .0001), and serrated polyps (CCE-WL, 57%; CCE-FICE, 74%; P = .0022). Both modes detected all adenocarcinomas. No significant differences were found between CCE-WL and CCE-FICE of the per-lesion sensitivity for ≥10-mm CRTs (CCE-WL, 81%; CCE-FICE, 94%; P = .1138) or protruding CRTs (CCE-WL, 77%; CCE-FICE, 86%; P = .0614). Kappa coefficients for overall CRTs for CCE-WL and CCE-FICE were .66 and .64, respectively, which indicated substantial agreement. CONCLUSIONS: CCE-FICE improved the detection rates for all CRTs except adenocarcinomas, ≥10-mm polyps, and protruding polyps when compared with CCE-WL. (Clinical trial registration number: UMIN 000021125.).
引用
- Omori T, Jodai Y, Maeda K, et al. Prospective study of diagnostic yields of flexible spectral imaging color enhancement installed in colon capsule endoscopy for colorectal polyps and tumors. Gastrointest Endosc. 2023. PMID: 37797727. doi:10.1016/j.gie.2023.09.002
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。