文献速递胰腺与胆道疾病消化内镜消化道肿瘤与早癌

全覆膜自膨式金属支架的锚定鳍影响拉脱力与支架移位

Anchoring fins of fully covered self-expandable metal stents affect pull-out force and stent migration.
PubMed 文献 · Gastrointest Endosc · 2023-10-18

本研究通过拉脱力测量量化全覆膜自膨式金属支架(FCSEMS)的抗移位能力,并回顾性分析单中心178枚FCSEMS治疗远端胆道狭窄的180天随访结局。带锚定鳍支架的拉脱力为14.2±0.1 N,高于喇叭口支架的3.8±1.0 N(P<.0001),临床移位率(5% vs 34%)及支架功能障碍相关不良事件更低。

要点

  • 带锚定鳍FCSEMS拉脱力14.2±0.1 N,高于喇叭口支架3.8±1.0 N(P<.0001)。
  • 带锚定鳍支架移位率5%,低于喇叭口支架34%(P<.0001)。
  • 带锚定鳍支架1、3、6个月通畅率分别为96%、90%、80%,均高于喇叭口支架。

核心结论

  • 拉脱力作为支架生物力学特性可预测FCSEMS在远端胆道狭窄中的抗移位能力,或可用于该类支架分类。
  • 带锚定鳍FCSEMS的移位率及不良事件率显著低于喇叭口FCSEMS。

研究设计与样本

  • 生物力学测量量化FCSEMS的拉脱力。
  • 单中心回顾性队列分析178枚FCSEMS,随访180天评估治疗成功与不良事件。

主要发现

  • 带锚定鳍支架最大拉脱力14.2±0.1 N,喇叭口支架3.8±1.0 N(P<.0001),前者约为后者4倍。
  • 带锚定鳍支架移位率5%,喇叭口支架34%(P<.0001)。
  • 因支架功能障碍行非计划ERCP:带锚定鳍组15%,喇叭口组29%(P=.046)。
  • 支架功能障碍所致胆管炎:带锚定鳍组5%,喇叭口组19%(P=.02)。
  • 1、3、6个月支架通畅率:带锚定鳍组96%、90%、80%,喇叭口组90%、74%、66%(log-rank检验P=.03)。

关键数值与推荐

  • 带锚定鳍FCSEMS最大拉脱力14.2±0.1 N。
  • 喇叭口FCSEMS最大拉脱力3.8±1.0 N(P<.0001)。
  • 带锚定鳍组移位率5%,喇叭口组34%(P<.0001)。
  • 带锚定鳍组非计划ERCP率15%,喇叭口组29%(P=.046)。
  • 带锚定鳍组胆管炎发生率5%,喇叭口组19%(P=.02)。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND AND AIMS: Stent migration and subsequent adverse events are frequently observed in the use of fully covered self-expandable metal stents (FCSEMSs) for distal biliary stenosis. In this study, we identified predictors for stent migration based on biomechanical stent characteristics and associated these findings with clinical outcomes. METHODS: The migration resistance of FCSEMSs was quantified by measuring the pull-out force. We analyzed a single-center retrospective cohort of 178 FCSEMSs for treatment success and adverse events occurring during 180 days of follow-up. RESULTS: Biomechanical measurements revealed a 4-fold higher migration resistance of FCSEMSs with anchoring fins (AF-FCSEMSs; Fmax = 14.2 ± .1 N) compared with FCSEMSs with flared ends (FE-FCSEMSs; Fmax = 3.8 ± 1.0 N; P < .0001). Clinically, AF-FCSEMSs showed lower rates of migration compared with FE-FCSEMSs (5% vs 34%, P < .0001). Unscheduled ERCP procedures because of stent dysfunction were less frequent in the AF group compared with the FE group (15% vs 29%, P = .046). Cholangitis because of stent dysfunction was observed in 5% of the AF group compared with 19% in the FE group (P = .02). Stent patency rates at 1, 3, and 6 months were higher in the AF group (96%, 90%, and 80%, respectively) compared with the FE group (90%, 74%, and 66%; log-rank test: P = .03). CONCLUSIONS: The pull-out force as a biomechanical stent property predicts the migration resistance of FCSEMSs in distal biliary stenosis and may thus be used to classify stents for this application. AF-FCSEMSs showed a significantly lower rate of migration and adverse events compared with FE-FCSEMSs.