文献速递衰弱、肌少症与跌倒老年综合评估

Gripwise与Jamar握力计在老年住院患者中的比较:新型握力计评估握力的挑战

Gripwise Versus Jamar: The Challenge of a New Dynamometer Assessing Handgrip Strength.
PubMed 文献 · J Gerontol A Biol Sci Med Sci · 2023-01-01

这项单中心横断面研究在65岁及以上住院患者中比较电子Gripwise与金标准Jamar握力计的最大值和均值相关性,组内相关系数分别为0.93和0.94。但Gripwise检出低握力值(女性<16 kg、男性<27 kg)的比例显著高于Jamar(71% vs 48%,p<.001),因此研究者确定了Gripwise的替代截断值(女性<12 kg、男性<22 kg)并在补充验证人群中验证,诊断性能较高。

要点

  • Gripwise与Jamar握力计在老年住院患者中相关性高(ICC 0.93-0.94)。
  • Gripwise记录的低握力值比例显著高于Jamar(71% vs 48%)。
  • Gripwise诊断低握力需采用替代截断值(女<12 kg,男<22 kg)。

核心结论

  • Gripwise与Jamar在老年住院患者中的相关性得到确认,但Gripwise记录的值更低,需采用替代截断值来诊断低握力。

研究设计与样本

  • 单中心横断面研究,纳入65岁及以上住院患者348例(平均年龄79±9岁,52%女性),随机分配先使用Jamar或Gripwise进行测量,可坐位或卧床评估。

主要发现

  • 两种设备最大值的组内相关系数为0.93(95%CI 0.92-0.94,p<.001),均值的组内相关系数为0.94(95%CI 0.93-0.95,p<.001)。
  • 以Jamar低握力标准(女性<16 kg,男性<27 kg)判定,Jamar检出48%患者低握力,Gripwise检出71%(p<.001)。
  • Gripwise替代截断值(女性<12 kg,男性<22 kg)在补充验证人群(n=70)中诊断性能高:女性灵敏度93%、特异度87%;男性灵敏度94%、特异度96%。

关键数值与推荐

  • Jamar低握力标准:女性<16 kg,男性<27 kg。
  • Gripwise替代低握力截断值:女性<12 kg,男性<22 kg。
  • Gripwise替代截断值的诊断性能:女性灵敏度93%、特异度87%;男性灵敏度94%、特异度96%。

临床怎么用

  • 在老年住院患者中使用Gripwise评估握力时,应采用替代截断值(女性<12 kg,男性<22 kg)诊断低握力,避免直接套用Jamar标准。

局限性

  • 本研究为单中心横断面设计,补充验证人群样本量较小(n=70)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Sarcopenia diagnosis is partly based on handgrip strength (HGS) assessment. The gold-standard dynamometer for this measurement is the Jamar. The electronic Gripwise is a smaller and lighter one, and its measurements are correlated with the Jamar's in laboratory tests. Our study aimed to confirm this correlation in aged patients. METHODS: This monocenter cross-sectional study was performed in patients of 65 years and older admitted at the University Hospital. Participants were assessed either in a seated or bedridden position, randomly allocated to begin the measurements with the Jamar or the Gripwise. RESULTS: Among 649 aged inpatients assessed for eligibility, 348 were included (mean age: 79 ± 9; 52% females). The intraclass correlation coefficient was 0.93 (95% confidence interval [CI] 0.92-0.94, p < .001) for the maximum value measured with both devices and 0.94 (95% CI 0.93-0.95, p < .001) for the mean values. However, there was a significant difference in detecting low values (<16 kg in women, <27 kg in men), found in 48% of patients with Jamar, and 71% with Gripwise (p < .001). Thus, we determined alternate cutoffs for diagnosing HGS low values with the Gripwise (<12 kg in women, <22 kg in men), further validated in a supplementary validation population (n = 70). The diagnostic performances of these alternative cutoffs were high (93% sensitivity and 87% specificity in women; 94% sensitivity and 96% specificity in men). CONCLUSIONS: The correlation of the Gripwise with the Jamar was confirmed in aged inpatients. However, lower values recorded with the Gripwise require alternate cutoffs for a relevant low HGS diagnosis.

来源:J Gerontol A Biol Sci Med Sci,查看原文。版权归原作者所有。

引用

  1. Villain C, Lebaube S, Kremer C, et al. Gripwise Versus Jamar: The Challenge of a New Dynamometer Assessing Handgrip Strength. J Gerontol A Biol Sci Med Sci. 2023. PMID: 37578949. doi:10.1093/gerona/glad198

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。