在腹腔镜腹壁疝修补计算机模型中测试补片
本研究基于工程学原理构建腹腔镜腹壁疝修补的计算机模型,用一名5 cm脐疝患者的CT扫描生成三维模型,在站立、咳嗽和跳跃对应的腹腔内压下测试四种补片(腹腔内铺置、重叠5 cm)。结果显示补片均显著降低腹壁应变(从无补片时的17.5%降至不足1%)并防止疝膨出,但不同补片在断裂力和弹性转为塑性时的力方面差异显著(p < 0.001)。
要点
- 补片在模型中均显著降低腹壁应变并防止疝膨出。
- 不同补片力学性能差异显著(p < 0.001)。
- 模型基于工程学原理,需进一步优化以模拟腹壁生物力学。
核心结论
- 所测试补片在计算机模型中均显著降低腹壁应变并防止疝膨出。
- 不同补片在断裂力和弹性转为塑性时的力方面差异显著(p < 0.001)。
研究设计与样本
- 基于工程学原理构建腹腔镜腹壁疝修补计算机模型。
- 用一名5 cm脐疝患者的CT扫描生成三维模型。
- 四种补片以腹腔内铺置、重叠5 cm应用,并在站立、咳嗽和跳跃的腹腔内压下测试。
- 对每种补片进行纵向和横向的机械失效测试。
主要发现
- 补片断裂力范围为14.8 (5.6) 至78 (5) n/cm,差异显著(p < 0.001)。
- 补片弹性转为塑性时的力范围为1.6 (0.1) 至14.2 (0.2) n/cm,差异显著(p < 0.001)。
- 无补片时腹壁应变为17.5%,加补片后降至不足1%。
- 所有补片均防止模型中的疝膨出。
临床怎么用
- 该模型可用于在术前比较不同补片的生物力学表现。
- 补片选择需考虑其断裂力和弹性转为塑性时的力等力学参数。
局限性
- 模型仅基于一名5 cm脐疝患者的CT扫描,可能不具普遍代表性。
- 模型需进一步优化以更好模拟腹壁生物力学。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The ideal mesh for hernia repair has yet to be found, in addition our knowledge of the biomechanics of the abdominal wall is poor. The aim of this study was to develop a computer model of a laparoscopic ventral hernia repair and to test different meshes in that model at various intra-abdominal pressures. METHODS: Four meshes were tested in a computer model of a ventral hernia. Mechanical failure testing of each mesh was performed in both the longitudinal and transverse directions. A CT scan of a patient with a 5 cm umbilical hernia was used to generate a 3 dimensional model. Meshes were then applied to the model in an intraperitoneal onlay position with a 5 cm overlap. The model was then tested with intraabdominal pressures for standing, coughing and jumping with and without meshes. RESULTS: Meshes varied significantly (p < 0.001) in both rupture force 14.8 (5.6) to 78 (5) n/cm and force in which they changed from elastic to plastic 1.6 (0.1) to 14.2 (0.2) n/cm. When applied to the computer model all significantly reduced the strain on the abdominal wall from 17.5% without mesh to less than 1% with mesh. All meshes prevented the hernia from bulging in the model. CONCLUSIONS: We have developed a computer model of laparoscopic ventral hernia repair based on engineering principles. This model demonstrated that meshes tested significantly reduced the strain on the abdominal wall. Further studies are required to refine this model in order to best simulate the biomechanics of the abdominal wall.
引用
- Qandeel H, Chew C, Tanner KE, et al. Testing meshes in a computer model of a laparoscopic ventral hernia repair. Surg Endosc. 2021. PMID: 34528129. doi:10.1007/s00464-021-08735-0
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。