新型Veress针去耦安全机制减少腹壁穿刺后过冲
本研究基于最小化设计思路开发了一种可减少Veress针(VN)过冲风险的安全机制,通过穿刺进入腹腔后立即将外科医生手部与针体去耦来防止针尖加速。在离体猪腹壁组织模型上,两名新手和一名有经验外科医生使用标准VN的过冲距离为41.4 mm [37.5-45 mm],而使用带去耦机制的VN+为20.8 mm [17.5-22.5 mm],差异有统计学意义(p < 0.001)。
要点
- 新型VN+去耦机制在离体猪腹壁模型中显著减少穿刺过冲。
- 标准VN过冲41.4 mm,VN+过冲20.8 mm,p < 0.001。
- pilot研究提示VN+至少减少50%过冲,需进一步临床验证。
核心结论
- 新型去耦安全机制成功整合入标准Veress针形成VN+,在标准化离体猪腹壁模型中至少减少50%的过冲。
研究设计与样本
- 基于最小化设计方法开发VN+去耦安全机制,并在离体猪腹壁组织模型上进行原型评估,由两名新手和一名有经验外科医生操作。
主要发现
- 标准VN的过冲距离为41.4 mm [37.5-45 mm],VN+为20.8 mm [17.5-22.5 mm],差异有统计学意义(p < 0.001)。
临床怎么用
- 该去耦机制可在穿刺进入腹腔后立即将外科医生手部与针体解耦,从而降低针尖过冲风险。
局限性
- 研究仅为离体猪腹壁模型上的初步试验,样本量有限,尚未在临床环境中验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Complications that occur in laparoscopic surgery are often associated with the initial entry into the peritoneal cavity. The literature reported incidences of Veress needle (VN) injuries of e.g. 0.31% and 0.23%. In a 2010 national survey of laparoscopic entry techniques in the Canadian General Surgical practice, 57.3% of respondents had either experienced or witnessed a serious laparoscopic entry complication like bowel perforation and vascular injury. As those complications are potentially life threatening and should be avoided at all costs, improving safety of this initial action is paramount. METHODS: Based on a bare minimum design approach with focus on function expansion of existing components, a new Safety mechanism was developed for the VN that decreases the risks of VN overshooting. The mechanism works by preventing the puncturing acceleration of the tip of the VN by decoupling the surgeon's hand from the VN immediately after entering the abdomen. RESULTS: Based on a set of requirements, a first prototype of the VN+ with force decoupling safety mechanism is presented and evaluated on an ex vivo porcine abdominal wall tissue model in a custom setup. The experiments conducted by two novices and one experienced surgeon indicated a significant difference between the attempts with a standard, conventional working VN (41.4 mm [37.5-45 mm]) and VN+ with decoupling mechanism (20.8 mm [17.5-22.5 mm]) of p < 0.001. CONCLUSION: A new decoupling safety mechanism was integrated successfully in a standard VN resulting in a VN+ . The results from the pilot study indicate that this new VN+ reduces overshooting with a minimum of 50% in a standardised ex vivo setting on fresh porcine abdominal wall specimens.
引用
- Postema RR, Cefai D, van Straten B, et al. A novel Veress needle mechanism that reduces overshooting after puncturing the abdominal wall. Surg Endosc. 2021. PMID: 34159463. doi:10.1007/s00464-021-08603-x
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。