文献速递甲状腺与颈部外科

经口内镜甲状腺切除术后神经监测:兼具剥离与刺激功能的内镜器械概念验证

Stimulating and dissecting instrument for transoral endoscopic thyroidectomy: proof of concept investigation.
PubMed 文献 · Surg Endosc · 2019-06-19

该前瞻性动物实验在12只小猪的24条喉返神经和24条迷走神经上,比较经皮单极刺激探头与两种内镜剥离器刺激模式的术中神经监测效果。结果显示内镜器械可行、安全,但诱发肌电反应所需电流高于经皮探头,尖端接触模式所需电流低于侧方接触模式。

要点

  • 内镜刺激剥离器在TOETVA动物模型中可行、安全,无相关并发症。
  • 内镜器械诱发迷走神经和喉返神经肌电反应所需电流高于经皮单极探头。
  • 剥离器尖端接触神经刺激所需电流低于侧方接触模式。

核心结论

  • 内镜刺激剥离器在TOETVA动物模型中可简单、有效、安全地监测迷走神经和喉返神经功能。
  • 内镜器械较经皮手探针需要更高电流才能诱发肌电反应,尖端接触模式较侧方模式所需电流更低。

研究设计与样本

  • 前瞻性实验研究,在12只小猪上实施TOETVA,共检测24条喉返神经和24条迷走神经。
  • 通过气管表面电极记录喉返神经肌电图,比较经皮单极刺激探头(I组)、内镜Maryland剥离器尖端接触(II组)和剥离器尖端侧方接触(III组)三种刺激方式。

主要发现

  • 所有TOETVA手术中内镜器械应用均可行,未导致任何并发症,24条喉返神经和迷走神经均被检测、刺激和监测。
  • 随刺激电流增加,肌电振幅增加,呈剂量-反应曲线。
  • 迷走神经平均刺激阈值:I组0.28 mA,II组0.56 mA,III组0.58 mA(P1=0.00,P2=0.00,P3=0.11)。
  • 诱发迷走神经最大反应的最小电流:I组0.65 mA,II组1.07 mA,III组1.14 mA(P1=0.00,P2=0.00,P3=0.48);诱发喉返神经最大反应的最小电流:I组0.6 mA,II组0.95 mA,III组1.05 mA(P1=0.00,P2=0.00,P3=0.31)。
  • 各组潜伏期值相似,重复(>10分钟)超最大(>5 mA)电刺激安全。

临床怎么用

  • 在TOETVA中可考虑使用兼具剥离与刺激功能的内镜器械进行迷走神经和喉返神经实时肌电监测。
  • 使用内镜器械刺激时需注意其诱发肌电反应所需电流高于经皮手探针,尖端接触模式较侧方模式所需电流更低。

局限性

  • 本研究为动物实验(猪模型),结论需进一步临床验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES/HYPOTHESIS: Intraoperative neuromonitoring (IONM) is a useful adjunct for recurrent laryngeal nerve (RLN) mapping and identification in transoral endoscopic thyroidectomy vestibular approach (TOETVA). This experimental study aimed to investigate the feasibility, safety, thresholds required of an endoscopic forceps that combine the function of surgical dissection and nerve stimulation. STUDY DESIGN: Prospective experimental research. METHODS: TOETVA was performed in 12 piglets, i.e., 24 RLNs and 24 vagal nerves (VN). RLNs electromyography (EMG) was recorded via endotracheal surface electrodes. Baseline EMG of VN and RLN were recorded and compared by (a) percutaneously placed monopolar stimulator probe (Group I), (b) adapted Maryland endoscopic dissector applied on nerves at its tip-end (Group II) and (c) endoscopic dissector tip-lateral applied (Group III). EMG profiles, amplitude, latency, waveform, thresholds and supra-maximal stimulation (5 mA) were analyzed. RESULTS: Application of the endoscopic device was feasible in all TOETVA and did not result in any morbidity. 24 RLNs and VNs were detected, stimulated and monitored. With increase of stimulation current, the amplitude of EMG increased, showing a dose-response curve. Mean VN stimulation thresholds were: Group I 0.28 mA, Group II 0.56 mA, Group III 0.58 mA (P1 = 0.00, P2 = 0.00, P3 = 0.11). Minimal current to evoked a maximal VN response was: Group I 0.65 mA, Group II 1.07 mA and Group III 1.14 mA (P1 = 0.00, P2 = 0.00, P3 = 0.48). Minimal current to evoke a RLN maximal response was Group I 0.6 mA, Group II 0.95 mA and Group III 1.05 mA (P1 = 0.00, P2 = 0.00, P3 = 0.31). Latency values were similar to each group. Repetitive (> 10 min) supra-maximal (> 5 mA) electrical stimulation was safe. CONCLUSIONS: The application of endoscopic stimulating dissector is simple, effective and safe way to monitor both VN and RLN function during a TOETVA animal model. It provides surgeons with real-time feedback of EMG response and can be applied as a tool for RLN monitoring. Endoscopic instrument required higher current to evoke EMG response compared to hand probe stimulation. Tip-end required less current to evoke EMG response compared to tip-lateral mode of stimulation.

来源:Surg Endosc,查看原文。版权归原作者所有。

引用

  1. Zhang D, Li S, Dionigi G, et al. Stimulating and dissecting instrument for transoral endoscopic thyroidectomy: proof of concept investigation. Surg Endosc. 2019. PMID: 31218426. doi:10.1007/s00464-019-06936-2

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。