近红外荧光叠加成像系统增强甲状旁腺可视化
本研究设计了一种叠加组织成像系统(OTIS),通过检测组织近红外自发荧光(NIRAF)并将信号以可见光直接背投到手术野,以增强甲状旁腺可视化。
要点
- OTIS将NIRAF信号背投至手术野,无需远程显示器或造影剂。
- OTIS离体测试对甲状旁腺、甲状腺和淋巴结的准确率为100%。
核心结论
- OTIS无需显示器或造影剂即可直接在手术野内增强甲状旁腺的无标记可视化,但需进一步优化设备以用于临床。
研究设计与样本
- 研究设计了一种叠加组织成像系统(OTIS),并在离体和活体(30例甲状腺切除和/或甲状旁腺切除患者)中测试其增强甲状旁腺可视化的能力。
主要发现
- OTIS在离体测试中对甲状旁腺、甲状腺和淋巴结的准确率为100%。
关键数值与推荐
- 系统准确性的判定依据为NIRAF投影与外科医生对原位甲状旁腺的视觉确认或切除标本的组织病理学报告之间的相关性。
临床怎么用
- OTIS可辅助外科医生在甲状腺或甲状旁腺手术中实时识别甲状旁腺,减少误判风险。
局限性
- 需进一步优化设备才能最终用于临床。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Misidentifying parathyroid glands (PGs) during thyroidectomies or parathyroidectomies could significantly increase postoperative morbidity. Imaging systems based on near infrared autofluorescence (NIRAF) detection can localize PGs with high accuracy. These devices, however, depict NIRAF images on remote display monitors, where images lack spatial context and comparability with actual surgical field of view. In this study, we designed an overlay tissue imaging system (OTIS) that detects tissue NIRAF and back-projects the collected signal as a visible image directly onto the surgical field of view instead of a display monitor, and tested its ability for enhancing parathyroid visualization. STUDY DESIGN: The OTIS was first calibrated with a fluorescent ink grid and initially tested with parathyroid, thyroid, and lymph node tissues ex vivo. For in vivo measurements, the surgeon's opinion on tissue of interest was first ascertained. After the surgeon looked away, the OTIS back-projected visible green light directly onto the tissue of interest, only if the device detected relatively high NIRAF as observed in PGs. System accuracy was determined by correlating NIRAF projection with surgeon's visual confirmation for in situ PGs or histopathology report for excised PGs. RESULTS: The OTIS yielded 100% accuracy when tested ex vivo with parathyroid, thyroid, and lymph node specimens. Subsequently, the device was evaluated in 30 patients who underwent thyroidectomy and/or parathyroidectomy. Ninety-seven percent of exposed tissue of interest was visualized correctly as PGs by the OTIS, without requiring display monitors or contrast agents. CONCLUSIONS: Although OTIS holds novel potential for enhancing label-free parathyroid visualization directly within the surgical field of view, additional device optimization is required for eventual clinical use.
引用
- McWade MA, Thomas G, Nguyen JQ, et al. Enhancing Parathyroid Gland Visualization Using a Near Infrared Fluorescence-Based Overlay Imaging System. J Am Coll Surg. 2019. PMID: 30769112. doi:10.1016/j.jamcollsurg.2019.01.017
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。