三种临床前染料近红外荧光腹腔镜输尿管成像的猪模型研究
该研究在3头雌性猪中评估了IRDye® 800BK、IRDye® 800NOS和IRDye® 800CW三种染料经静脉注射后行近红外荧光腹腔镜输尿管显像的效果,每种染料6 mg静脉注射。结果显示三种染料均能识别输尿管解剖走行,其中IRDye® 800BK的目标-背景比最高(2.14)。离体实验显示荧光强度与染料浓度及镜头距离相关。
要点
- 三种临床前染料经静脉注射均可实现猪活体输尿管近红外荧光显像
- IRDye® 800BK的输尿管荧光出现时间35分钟,目标-背景比最高为2.14
- 离体实验中三种染料在各浓度及1、5、10 cm距离均产生强荧光信号
核心结论
- 静脉注射IRDye® 800CW、IRDye® 800BK和IRDye® 800NOS均有助于在活体猪模型中识别输尿管的解剖走行。
- 测得的最高目标-背景比出现在使用IRDye® 800BK时。
研究设计与样本
- 在3头雌性荷兰长白猪中评估三种临床前染料,每头猪测试1种染料,静脉注射剂量6 mg,浓度1 mg/ml。
- 使用腹腔镜成像系统在荧光模式和白光模式下进行成像。
- 另进行离体成像实验,每种染料以PBS稀释为8个递减浓度,在透明试管中于距腹腔镜1、5和10 cm处行近红外荧光模式评估。
主要发现
- 三种染料均能通过近红外荧光成像识别输尿管。
- IRDye® 800BK、IRDye® 800NOS和IRDye® 800CW使输尿管出现荧光的时间分别为35、45和10分钟。
- IRTye® 800BK、IRDye® 800NOS和IRDye® 800CW的最大目标-背景比分别为2.14、0.66和1.44。
- 离体实验中,三种染料在所有评估浓度和所有评估距离下均产生强荧光信号。
- 离体观察到荧光信号强度与染料浓度及与物体的距离之间存在相关性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Ureteric injury is reported to occur in 1-7.6% of colorectal surgeries. To reduce the incidence of ureteral injury, it is essential to identify the ureters. The use of near-infrared fluorescence (NIRF) imaging with intravenously administered dyes might be of added value for ureteral visualization during laparoscopy. The aim of this study is to assess the performance of three preclinical dyes; IRDye® 800BK, IRDye® 800NOS and IRDye® 800CW, for near-infrared fluorescence laparoscopy of the ureter in pigs. METHODS: In three female Dutch landrace pigs, the new dyes were evaluated. In each pig, 1 dye was tested using a 6-mg intravenous dose in a concentration of 1 mg/ml. Imaging was performed in fluorescence mode and white light mode with a laparoscopic imaging system. In order to further evaluate the dyes, an ex vivo imaging experiment was performed, in which 8 decreasing concentrations per dye, diluted in PBS, were evaluated in a transparent test tube with NIRF mode at a distance of 1, 5 and 10 cm from the laparoscope. RESULTS: All three dyes were effective in allowing the identification of the ureter with NIRF imaging. The ureter became fluorescent after 35, 45 and 10 min, respectively, for IRDye® 800BK, IRDye® 800NOS and IRDye® 800CW with a maximum target-to-background ratio (TBR) of 2.14, 0.66 and 1.44, respectively. In the ex vivo imaging experiment, all three dyes produced a strong fluorescence signal at all concentrations and all distances evaluated. CONCLUSIONS: Intravenous administration of the preclinical dyes IRDye® 800CW, IRDye® 800 BK and IRDye® 800NOS facilitated successful identification of the anatomical course of the ureter in living pig models. The highest measured TBR occurred with the use of IRDye® 800BK. Ex vivo, a correlation was observed between the fluorescence intensities of the signal with the concentration of the dye and with the distance to the object.
引用
- van den Bos J, Al-Taher M, Bouvy ND, et al. Near-infrared fluorescence laparoscopy of the ureter with three preclinical dyes in a pig model. Surg Endosc. 2018. PMID: 30478696. doi:10.1007/s00464-018-6596-z
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。