警示:切勿使用钉枪将补片固定于膈肌
制造商说明书和现有指南均禁忌在膈肌使用钉枪,推荐替代固定技术。结论强调不应在肋弓以上的膈肌使用钉枪固定补片。
要点
- 23/25例(92%)心脏损伤由膈肌钉枪触发。
- 制造商说明书和指南均禁忌膈肌使用钉枪。
核心结论
- 不应在肋弓以上的膈肌使用钉枪进行补片固定。
研究设计与样本
- 系统文献检索(2017年5月)识别25例严重穿透性心脏并发症,并回顾制造商说明书。
主要发现
- 25例中23例(92%)严重心脏损伤及心脏压塞由膈肌使用钉枪触发。
- 6例(24%)涉及延伸至剑突下区域的腹壁和切口疝,19例(76%)涉及补片加强的食管裂孔疝修补。
关键数值与推荐
- 制造商说明书禁忌在膈肌尤其是心脏附近部署钉枪。
- 现有指南紧急建议避免在膈肌使用钉枪,推荐替代固定技术。
临床怎么用
- 在膈肌(尤其肋弓以上)固定补片时,应避免使用钉枪,选择替代固定技术。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The mesh fixation technique used in repair of hiatal hernias and subxiphoid ventral and incisional hernias must meet strenuous requirements. In the literature, there are reports of life-threatening complications with cardiac tamponade and a high mortality rate on using tacks. The continuing practice of tack deployment for mesh fixation to the diaphragm and esophageal hiatus should be critically reviewed. METHODS: In a systematic search of the available literature in May 2017, 23 cases of severe penetrating cardiac complications were identified. The authors became aware of two other cases in which they acted as medical experts. Furthermore, the instructions for use issued by the manufacturers of the tacks were reviewed with regard to their deployment in the diaphragm. RESULTS: Twenty-three of 25 cases (92%) with severe cardiac injuries and subsequent cardiac tamponade were triggered by the use of tacks in the diaphragm. In six cases (24%), these related to ventral and incisional hernias with extension to the subxiphoid area, and in 19 cases (76%) to mesh-augmented hiatoplasty. Twelve of 25 (48%) patients died as a result of pericardial and/or heart muscle injury with cardiac tamponade despite heart surgery intervention. In the tack manufacturers' instructions for use, their deployment in the diaphragm, in particular in the vicinity of the heart, is contraindicated. Likewise, the existing guidelines urgently advise against the use of tacks in the diaphragm, recommending instead alternative fixation techniques. CONCLUSIONS: Tacks should not be used for mesh fixation in the diaphragm above the costal arch.
引用
- Köckerling F, Schug-Pass C, Bittner R, et al. A word of caution: never use tacks for mesh fixation to the diaphragm! Surg Endosc. 2018. PMID: 29340811. doi:10.1007/s00464-018-6050-2
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。