文献速递胃肠外科急腹症与外科感染

新型单步全层切除装置安全成功切除困难胃肠病变

Safe and successful resection of difficult GI lesions using a novel single-step full-thickness resection device (FTRD®).
PubMed 文献 · Surg Endosc · 2017-06-29

本研究纳入60例患者,评估新型全层切除装置(FTRD®)行内镜下全层切除(eFTR)的安全性、学习曲线、R0切除率及临床结局。技术成功率达97%(58/60),组织学R0切除率79%,不良事件发生率7%,无继发穿孔或eFTR相关死亡。结论认为经特定培训后eFTR可行、安全且有前景,但需更大规模对照研究证实。

要点

  • FTRD®行eFTR技术成功率达97%(58/60),可到达标记病变并完成切除。
  • 组织学R0切除率79%,全层切除率88%,基于随访组织学的临床成功率达88%。
  • 不良事件发生率7%,无继发穿孔或eFTR相关死亡,仅1例(2%)需术后手术。

核心结论

  • 经特定培训后,eFTR是可行、安全且有前景的一体化内镜下切除技术。
  • eFTR可完整切除累及黏膜下层的胃肠壁病变,不良事件风险低。
  • 初步结果需在更大规模对照研究中证实。

研究设计与样本

  • 单中心病例系列,60例患者在本机构接受eFTR。
  • 评估安全性、学习曲线、R0切除率和临床结局。

主要发现

  • 97%(58/60)的干预中FTRD®装载内镜到达预先标记病变并完成eFTR(技术成功)。
  • 88%的病例实现全层切除,组织学检查R0切除率为79%。
  • 基于随访组织学的临床成功率更高,达88%。
  • 不良事件发生率为7%。
  • 残余阑尾腺瘤eFTR后阑尾炎导致唯一一例术后手术(1/58,2%)。

临床怎么用

  • 对于向肠壁深层生长或位于易穿孔解剖部位的病变,eFTR可作为经典内镜下切除(EMR和ESD)的替代选择。
  • 实施eFTR前需经过特定培训。
  • eFTR后需关注残余阑尾炎等不良事件,多数轻微出血可内镜处理。

局限性

  • 初步结果需在更大规模对照研究中证实。
  • 原文未报告学习曲线的具体数值。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND AND STUDY AIMS: Classic endoscopic resection techniques (EMR and ESD) are limited to mucosal lesions. In the case of deeper growth into the gut wall and anatomic sites prone to perforation, the novel full-thickness resection device (FTRD®) opens a new dimension of possibilities for endoscopic resection. PATIENTS AND METHODS: Sixty patients underwent endoscopic full-thickness resection (eFTR) at our institution. Safety, learning curve, R0 resection rate, and clinical outcome were studied. RESULTS: In 97% (58/60) of the interventions, the FTRD®-mounted endoscope reached the previously marked lesion and eFTR was performed (technical success). Full-thickness resection was achieved in 88% of the cases, with an R0 resection on histological examination in 79%. The clinical success rate based on follow-up histology was even higher (88%). Adverse events occurred in 7%. Appendicitis of the residual cecal appendix after eFTR of an adenoma arising in the appendix led to the only post-eFTR surgery (1/58, 2%). Minor bleeding at the eFTR site (2/58, 3%) and an eFTR performed accidently without proper prior deployment of the OTSC® (1/58, 2%) were successfully treated endoscopically. There was no secondary perforation or eFTR-associated mortality. CONCLUSIONS: After specific training, eFTR is a feasible, safe, and promising all-in-one endoscopic resection technique. Our data show that eFTR allows complete resection of lesions affecting layers of the gut wall beneath the mucosa with a low risk of adverse events. However, our preliminary results need to be confirmed in larger, controlled studies.

来源:Surg Endosc,查看原文。版权归原作者所有。

引用

  1. Valli PV, Mertens J, Bauerfeind P, et al. Safe and successful resection of difficult GI lesions using a novel single-step full-thickness resection device (FTRD®). Surg Endosc. 2017. PMID: 28664442. doi:10.1007/s00464-017-5676-9

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。