带IC标签的内镜标记夹与接收天线用于腹腔镜手术定位检测
研究开发了一种带集成电路(IC)标签的内镜标记夹及接收天线,用于腹腔镜手术中客观定位。在猪腹腔镜手术模型中成功体内检测到标记夹,并在5例人结肠组织离体样本中全部(100%)准确检测。该技术有望替代传统印度墨纹身标记,避免墨水扩散导致腹膜炎的风险。
要点
- 新型IC标签标记夹可在猪腹腔镜模型中经肠壁被天线检测到
- 离体人结肠组织5例全部成功检测,成功率100%
- 该技术旨在避免传统印度墨纹身标记的腹膜炎风险
核心结论
- 带IC标签的内镜标记夹联合接收天线可在腹腔镜手术中实现客观定位,在猪体内模型和人离体结肠组织中均成功检测。
研究设计与样本
- 研究在猪腹腔镜手术模型中进行体内放置与检测,并采用5例人结肠组织离体样本评估检测能力。
主要发现
- 在猪腹腔镜手术模型中,通过腹腔内接收天线成功检测到置于胃肠道的IC标签标记夹。
- 在5例人结肠组织离体样本中,接收天线靠近浆膜面时全部(5/5,100%)准确检测到黏膜表面的IC标签标记夹。
临床怎么用
- 该技术尚处于动物实验和离体组织验证阶段,尚未报告临床应用数据。
局限性
- 研究仅基于猪模型和离体人结肠组织,样本量有限,未在临床患者中验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Conventional tattooing using India ink for preoperative marking during colonoscopy has been widely used. However, the ink used in this method can spread into the abdominal cavity and induce peritonitis. Therefore, we developed a new marking method using an endoscopic clip with an integrated circuit (IC) tag to accurately identify an objective location. We applied the novel method in laparoscopic surgery using a porcine model and also evaluated it with resected human gastrointestinal tissue. METHODS: We placed an endoscopic clip with an IC tag through a forceps aperture in the gastrointestinal tract, by using a porcine surgery model. After the endoscopic procedure, we performed laparoscopic surgery and approached the receiving antenna in the abdominal cavity to detect the IC tag through the porcine intestine. In an ex vivo examination with human colon tissues, the clip with the IC tag was placed on the mucosal surface to determine its ability to be detected. The receiving antenna near the serosal side of the human colon segment accurately detected the clip with the IC tag. RESULTS: We detected the clip with an IC tag with a detection device, by using a laparoscopic surgery model in vivo. We also confirmed its usefulness in five of five (100 %) human colon tissue samples tested ex vivo. CONCLUSION: We developed a novel marking device using an IC tag to identify an objective location. We successfully demonstrated the usefulness of the clip with the IC tag and the antenna device used for its detection in a porcin laparoscopic surgery model and in resected human colon tissue.
引用
- Wada Y, Miyoshi N, Ohue M, et al. Endoscopic marking clip with an IC tag and receiving antenna to detect localization during laparoscopic surgery. Surg Endosc. 2016. PMID: 27800589. doi:10.1007/s00464-016-5303-1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。